and press wheel to proceed to next step.
Note: buffer solution value may change if environment temperature it’s different than 20°C. Read solution’s label for more
information. According to this occurrence “pH Default” must be changed.
Comp Auto / Select Temp.
Once into submenu, to enable automatic temperature compensation, move wheel on “DISABLE” , press it and change
option to “ENABLE”. This procedure will automatically set temperature compensation.
Otherwise exit from this menu, move wheel on “Select Temp” and according to the following table enter required tempe
-
rature. This procedure will manually set temperature compensation.
End procedure by moving cursor on “Exit” from “Calibration pH” main menu and press it. If an error occurred during ca
-
libration procedure then the instrument will show an error message and will ask to proceed to a new calibration, cancel
current operation or restore default settings.
fig. 13