Series 

IZS31

 

Specific Product Precautions 4

 

Be  sure to  read  this before handling.

 

Back  page 4 

 

 

 
 
 
 

    

Operating

 

Environment

 

/

 

Storage

 

Environment

  

 

 

Warning

 

1. Observe the  operating fluid  temperature and 

ambient temperature 

range.

 

Fluid  and   ambient temperature  ranges  are   0  to  50

°

for  the 

ionizer,  feedback sensor  and   autobalance sensor.  Do  not  use 
the  ionizer  in locations subject to sudden temperature changes 
even   if  the  ambient temperature  range  is  within  the  specified 
limits, as condensation may  result. 

2.  Do  not  use this  product in  an  enclosed space.

 

This  product   utilizes  a  corona discharge  phenomenon.  Do  not 
use   the  product   in  an  enclosed space  as  ozone and   nitrogen 
o

xides   exist   in   such   places,   even    though    in   marginal 

quantities.

 

3. Environments to  avoid

 

A

void  using  and    storing  this  product    in  the  following 

environments 

since  they  may  cause damage to  this  product. 

a)  Avoid using  in a  place  that  exceeds an  ambient 

temperature range 

of 0  to 50 

°

C.

 

b)  Avoid using  in a  place  that  exceeds an  ambient  humidity 

range 

of 35  to 80  % Rh. 

c)  Avoid using  in a  place  where  condensation occurs due  to a 

drastic  temperature change. 

d)  Avoid using  in a  place  in the  presence of corrosive or 

explosive 

gas  or where  there  is  a  volatile  combustible. 

e)  Avoid using  in an  atmosphere where  there  are  particles, 

conductive 

iron powders, oil mist,  salt,  solvent,  blown  dust, 

cutting  oil (water,  liquid), etc. 

f)   Avoid using  in a  place  where  ventilated air  from an  air 

conditioner 

is directly  applied  to the  product. 

g)  Avoid using  in a  closed place  without  ventilation. 

h)  Avoid using  in direct  sunlight  or radiated heat. 

I)   Avoid using  in a  place  where  there  is a  strong  magnetic 

noise  (strong  electric  field, strong  magnetic field, or surge). 

j)   Avoid using  in a  place  where  static  electricity  is  discharged 

to the  main  body. 

k)  Avoid using  in a  place  where  a  strong  high  frequency 

occurs.

 

l)   Avoid using  in a  place  where  this  product  is  likely to be 

damaged 

by  lightning. 

m) Avoid using  in a  place  where  direct  vibration  or shock  is 

applied  to  the  main  body. 

n)  Avoid using  in a  place  where  there  is a  force  large  enough 

to deform  this  product  or weight  is applied  to  the  product. 

4. Do  not  use an  air containing mist  or  dust.

 

The  air  containing  mist  or  dust   will cause  the  performance  to 
decrease 

and  shorten the  maintenance  cycle. 

Suppl

y  clean  compressed  air  by   using  an   air  dryer   (IDFA 

series), 

air  filter (AF/AFF  series), and  mist  separator (AFM/AM 

series)

 

5. The  ionizer  and  sensors are  not  protected against 

a  surge caused by  a  lightning.

 

                                   

Ma

intenance

  

 

 

Warning

 

1.  Periodically  (for example, every  two  weeks)   inspect 

the  ionizer  and  clean  the  electrode needles.

 

Conduct 

a   regular   maintenance  to  see  if   the   product   is  run 

having  a  disorder. 
Maintenance 

should   be   conducted  by  a   fully  knowledgeable 

and  experienced person about  the  equipment. 
Using  for  a  long  period   of  time  will lower  the  static   electricity 
eliminating   performance,  if   particles  attach  to   the   electrode 
pin.

 

Replace 

the  electrode  cartridge, if  the  pins  are  rough   and   the 

static   electricity  eliminating   performance does  not  return   even 
after  being  cleaned. 

 

Danger  High  Voltage!

 

This  product    contains  a  high  voltage  generation  circuit.  When 

performing  maintenance  inspection,  be   sure  to   confirm  that   the 

power   supply  to   the   ionizer  is  turned   off.  Never   disassemble  or 

modify  the  ionizer,  as  this  may  not  only  impair  the  product’s 

functionality  but  could  cause an  electric  shock  or  electric  leakage. 

 

2.  When  cleaning  the  electrode  pin  or  replacing the 

electrode  cartridge, be  sure   to  turn  off  the  power 
supply 

to  the  main  body.

 

Touching   an  electrode needle  when   it is  electrified  may  result 
in electric  shock  or other  accidents. 

3.  Do  not  disassemble or  modify  this  product.

 

O

therwise,  an   electrical  shock,  damage  and/or   a   fire  may 

occur.   Also,   the   disassembled  or   modify   products  may   not 
achieve 

the   performances  guaranteed  in  the   specifications, 

and    excercise   caution   because   the    product    will   not   be 
warrantied.

 

 

 

                                       

Handling

  

 

 

Warning

 

1.  Do  not  drop,  bump  or apply  excessive impact  (10  G

 

or more)  while  handling.

 

Even   though   it  does  not  appear  to  be  damaged,  the   internal 
parts  may  be  damaged and  cause a  malfunction. 

2. When  mounting/dismounting  the  cable,  use  your 

finger  to  pinch  the  claw  of  the  modular  plug,  then 
attach/detach it correctly.

 

If 

the  modular   plug  is  at  a  difficult angle   to  attach/detach,  the 

modular   jack's   mounting   section  may  be  damaged  and   cause 
a  disorder. 

3. Do   not   operate   this   product   with   wet   hands.

 

Otherwise,  an   electrical  shock  or  accident   may 
occur.

 

Summary of Contents for IZS31 Series

Page 1: ...ries IZS31 height of sensor 10 mm sensor 1000 1500 2000 mm sensors are available electricity by a feedback sensor accordance with the detected polarity of a w min ANR per nozzle Conditions Static elec...

Page 2: ......

Page 3: ...utobalance sensor Autobalance sensor Measures the ion balance condition balance adjustment and reduction in ion hours by using an autobalance sensor Time n autobalance balancing input installation or...

Page 4: ...edback electricity removal has been increased by reading the the feedback sensor and then continuously emitting Supply pressure 0 1 MPa 7 l Installation height of 20 Ionizer Without 15 With sensor 10...

Page 5: ...edback sensor is used By outputting PLC etc it is possible to control the Ionizer Feedback sensor surface short static varies depending Charged object Enhanced display functions 1 Visualisation of cha...

Page 6: ...n Ion balance 30 V Applicable to environments sensitive to metal contamination Stainless steel Ion balance Made to Order Non standard bar length IZS31 X10 Standard 300 380 620 780 1100 1260 X10 460 54...

Page 7: ...liminating static electricity Prevents adhesion of dust Prevents winding failure due to wrinkles e e ed d d goods die Eliminating static electricity on fil Prevents attaching to conveyer Prevents disp...

Page 8: ......

Page 9: ...llation distance Discharge time sec Discharge time sec Note Static electricity elimination features are based on data from using a charged plate size 150 mm x 150 mm capacitance ANSI ESD STM3 1 2000 U...

Page 10: ...stallation height of feedback sensor 100 mm 600 mm Ion balance V Front Rear distance to the ionizer Charged object location mm Technical Data 2 Note Static electricity elimination features are based o...

Page 11: ...m x 150 mm capacitance ANSI ESD STM3 1 2000 Use this as a guideline for model selection only because the value varies depending on the TU0425 Length 10 mm KQ2T04 00 KQ2T04 b Both sides air supply IZS3...

Page 12: ...0 380 620 780 With 2 pcs 1100 1260 1500 With 1900 2300 With Series IZS31 How to Order 780 Made to Order Refer to the below table needle material Sensor Output Bracket End bracket Center bracket Power...

Page 13: ...low depends Two end brackets are always required regardless of the bar Bar length mm Quantity End bracket Center 300 380 620 780 2 pcs 1100 1260 1500 1900 2300 Ionizer Series Autobalance sensor IZS31...

Page 14: ...or ion balance adjustment of required security covers IZS31 E4 IZS31 E5 1 1 1 1 1 3 2 2 5 2 4 electrode cartridges 3 4 5 security covers Mounted part of 80 L indicate that the main unit is to be shipp...

Page 15: ...DC DC 7000 V 30 V Stainless electrode needle 100 V Air Clean and dry 0 7 MPa or less 4 24 VDC 10 200 mA or less While standing by 120 mA or less 200 mA or less When sensor is not used 170 mA or less 1...

Page 16: ...ricity from nonconductive workpieces Continuous Static Electricity Removal mode is recommended Note When the feedback sensor is installed at a height of 25 mm 2 Pulse DC mode Alternatively emits posit...

Page 17: ...he electrode be cleaned due to such deterioration a maintenance display be detected by only connecting a feedback sensor or autobalance sensor start signal description i u y t r e w q Type Contents Gr...

Page 18: ...rring to Static electricity Removal Characteristics main unit can also be used at other distances it may fail to operate verify that the main unit is functioning normally hole facing the charged surfa...

Page 19: ...es Connection needs Contents Ionizer driving power supply cable Outputs the workpiece s electrostatic potential as an analogue signal 1 to 5 V Settings 1 Sensing DC Mode supply the terminal is not gro...

Page 20: ...icates abnormal states of LED name ALARM HV Illuminates when an abnormal SNSR Illuminates when the feedback NDL CHECK Illuminates when electrode Blinks while the contamination Settings 1 Sensing DC Mo...

Page 21: ...display comes on and a maintenance output signal is output Turn off the power clean the electrode needles and turn the power on again 10 Timing chart Timing chart during normal operation Electric char...

Page 22: ...frequency is When the ion generation frequency is changed adjust the ion balance r adjustment the autobalance sensor may be removed as io adjustment will not be performed again until a maintenance sta...

Page 23: ...de Exercise caution 7 Air piping For single side piping block the unused port with Ionizer Series Switch setting 0 1 2 3 4 5 6 7 Connection needs Contents Ionizer driving power supply cable Settings 2...

Page 24: ...abnormal states of LED name ALARM HV Illuminates when an abnormal SNSR Illuminates when the autobalance NDL CHECK Illuminates when electrode Blinks when the stain Settings 2 Pulse DC Mode input and s...

Page 25: ...ON OFF ON OFF Operation permitted Power supply 24 VDC Discharge stop signal Input Input ON OFF ON OFF Operation permitted Static electricity removal ON Electrode needle stain detection or ion balance...

Page 26: ...signal Minimum wiring requirement for ionizer operation Wiring necessary to use various functions Wiring not required in the sensing DC mode Exercise caution 5 Air piping For single side piping block...

Page 27: ...harge stop signal Input Input ON OFF ON Operation permitted OFF State of emitting ions ON OFF Emission Ionizer Series Settings 3 DC Mode power input and sensor connection Function power is supplied Gr...

Page 28: ...izer driving power supply GND 24 VDC 10 Input Output signal power supply GND 24 VDC 10 PLC OUTPUT or 24 V Isolation circuit Photo coupler Gray IN2 Isolation circuit Photo coupler Gray IN2 or OUTPUT Is...

Page 29: ...80 x n 1 70 25 n No L Dimension L 14 16 36 19 14 KJH04 cartridge Maintenance level selection switch Frequency selection switch Ion balance adjustment trimmer IZS31 Ionizer full length L 2 31 33 31 22...

Page 30: ...ensor IZS31 DG 12 71 6 1 Detection part 2 x 3 4 Head 95 98 Power supply cable IZS31 CP 10 2 4 8 10 6 6 4 8 10 2 6 2 15 10 2 7 7 15 7 7 8 8 8 8 8 23 4 8 10 2 13 14 8 23 4 8 10 2 13 14 60 6 5 7 8 part 2...

Page 31: ...52 Order 1 specifications and delivery time please contact SMC compliant For 80 x n 1 70 L 14 KJH04 M5 X34 16 36 19 14 n No of electrode L Dimension Weight Part no cartridge Maintenance level selectio...

Page 32: ...12 is together Fitting M 5P X112 Note KJH04 M5 X34 Part no IZS31 300 IZS31 380 IZS31 620 IZS31 780 IZS31 1100 IZS31 1260 20 16 58 17 Order 2 specifications and delivery time please contact SMC electro...

Page 33: ...e when configuring a system 2 Only trained personnel should operate pneumatically operated machinery and equipment Since the product generates a high voltage handling it incorrectly can be dangerous A...

Page 34: ......

Page 35: ...e an enough space for maintenance piping and wiring Please take into consideration that the one touch fittings for supplying air need enough space for the air tubing to be easily attached detached To...

Page 36: ...he effects of Brown from the wall as tatic potentials within the clearances Blue Class D ground object 3 Be sure to turn off the powe wiring including attachment detachment connector 4 To connect a fe...

Page 37: ......

Page 38: ...t 4 Do not use an air containing mist or dust The air containing mist or dust will cause the performance to decrease and shorten the maintenance cycle Supply clean compressed air by using an air dryer...

Page 39: ...equired to supply air to the ionizer only for providing an opportunity to decrease energy saving countermeasure diagram 3 4 5 6 7 8 Air Filter Series AF Eliminates solid foreign matters such as power...

Reviews:

Related manuals for IZS31 Series