8.2. RDS 200N
The RDS 200N in combination with the AutoWave as an arbitrary generator can be used to test according to Ford
ES-XW7T, test procedure CI 230, simultaneously onto a max 4 channel DUT. The RDS 200N includes a sink and
can therefore also be used to generate arbitrary waves as required in different standards. It is also used as an
external power supply needed for voltage dip testing.
The RDS 200N can only be used in combination with other generators of the series 200 or with an external signal
source.
8.2.1. Frontside RDS 200N
1
Display
2
Voltage / Current LED
3
OV LED overvoltage ( analog input)
4
OT LED overtemperature
5
EUT Test supply output
1
Display
Display of the output voltage and current
2
Voltage / Current LED
The color of the LED displays the following status
green Voltage control
red
Current limiter is active, the output voltage is reduced.
3
LED OV (overvoltage)
The LED indicates an overvoltage of the analog input ( 0-10V DC)
4
LED OT (overtemperature)
The LED indicates an overtemperature of the device. The output voltage will be switched off.
5
EUT test supply
The EUT is powered via the safety laboratory plugs at the front panel of the simulator.
The maximum current at this output should not be greater than 16A.
6
GND connector
Ground connector for connect the reference earth.