Problem
Possible Cause
Measure
General
Flow reading with no flow present
System may not have been zeroed.
1.
Zero the system.
2.
While doing so, ensure that the tube is
filled completely but the liquid must not
move.
If the problem persists, please return the
sensor and the BioProTT
TM
FlowTrack SL for
servicing.
BioProTT
TM
FlowTrack SL gets very
warm
The BioProTT
TM
FlowTrack SL might be
exposed to an external heat source/
direct sunlight.
The BioProTT
TM
FlowTrack SL might
be covered so that heat cannot be
dissipated.
1.
Check if there is enough space around
the BioProTT
TM
FlowTrack SL for the heat
to dissipate.
2.
Ensure that the device is not exposed to
direct heat and/or sunlight.
If the problem persists, please return the
sensor and the BioProTT
TM
FlowTrack SL for
servicing.
Display
Display is blank frozen.
Power supply may be disconnected.
1.
Check if the device is connected to the
power supply.
2.
Ensure that the connection is tight.
3.
Power the device off.
4.
Power the device on again.
If the problem persists, please return the
sensor and the BioProTT
TM
FlowTrack SL for
servicing.
Missing information/values on
the screen
Display may have frozen.
1.
Refresh the screen by switiching to
another screen and back again.
2.
If no flow value is displayed, check if the
coupling is withing the specified range
of >50 %.
If the problem persists, please return the
sensor and the BioProTT
TM
FlowTrack SL for
servicing.
BioProTT™ FlowTrack SL
Page 46
Copyright em-tec GmbH
D101-704 BioProTT
TM
FlowTrack SL – User Manual – V 1.0