Command
Meaning
T1 <CR>....T7 <CR>
Request to set/use calibration table number 1...7
Please note:
If requesting a non-existing table (e.g. selecting table number 7 if there
are only five tables stored) the table will not change.
Cx.yz <CR>
set calibration factor (values between 0.50 and 1.50 with two decimal digits, e.g.
C1.10<CR>)
Table 14: Commands for digital interface of the BioProTT
TM
FlowTrack SL
The digital interface transmits the following ASCII data string every 100 ms (10 Hz):
ee_ss_aaa_p.pp_fffffff_mmmmmmm_sssssss_tttttt_<CR><LF>
Meaning of the Different Bytes:
Byte
Meaning
Remark
ee
hexadecimal error message according to separate table in
chapter 14 "
Error Code Information for the BioProTT
"
00 means no error is present
"_"
blank delimiter between the different parameters (hex 0x20)
Underscore is only used
to represent blank
ss
hexadecimal status information
see chapter 11 "
" for more details
aaa
acoustic coupling/RSS in %; range 0 to 100%
p.pp
calibration factor
ffffff
mean flow calculated over 100 ms (output rate of 10 Hz) in ASCII
(range: -999999 ml/min to +999999 ml/min) including calibration
factor
Note*
mmmmmm
mean flow calculated over 1 s (output rate of 1 Hz) in ASCII
(range: -999999 ml/min to +999999 ml/min) with calibration
factor applied
Note*
ssssss
mean flow calculated over 10 s (output rate of 0.1 Hz) in ASCII
(range: -999999 ml/min to +999999 ml/min) including calibration
factor
Note*
tttttt
board temperature in °C (38 = 38°C)
<CR>
0x0d Carriage Return
<LF>
0x0a Line Feed
Table 15: Output of digital interface of the BioProTT
TM
FlowTrack SL
*
Note:
Beside numerical values, these strings contain the following characters:
•
"-", a minus symbol is used for negative values or to indicate blanked out values (character hex "0x2D")
•
"^", indicating overflow condition (character hex: "0x5E")
•
"v"; indicating underflow condition (character hex: 0x76")
BioProTT™ FlowTrack SL
Page 32
Copyright em-tec GmbH
D101-704 BioProTT
TM
FlowTrack SL – User Manual – V 1.0