14 Error Codes and Information
Description
Digital Interface RS-232
Error Byte "ee"
Display
No error
00
flow value
RSS value
FLASH error during start-up test
01
error page w/01
FLASH error during runtime test
02
error page w/02
RAM error during runtime RAM test
08
error page w/08
Error in task handling
0F
error page w/0F
EEPROM test error (sensor)
19
error page w/19
Board temperature error (>75°C)
1A
error page w/1A
+15 V check error
1B
error page w/1B
-15 V check error
1C
error page w/1C
EEPROM CRC error (sensor)
21
error page w/21
EEPROM factor inconsistent (sensor)
22
error page w/22
EEPROM of the sensosor not
consistent (sensor)
27
error page w/27
Complementary values in EEPROM
not equal (sensor)
29
error page w/29
Illegal sensor type or sensor version
2A
error page w/2A
Sensor parameter not correct
2B
error page w/2B
Wrong CRC in burst table
2E
error page w/2E
Table 20: Error Information on the Different Interfaces
BioProTT™ FlowTrack SL
Page 44
Copyright em-tec GmbH
D101-704 BioProTT
TM
FlowTrack SL – User Manual – V 1.0