Pre-installation guide
51
7 Pre-installation guide
Following requirements apply, when installing the analyzer:
Carrier gas
Oxygen 99.95 % pure; 2 - 4bar (30 - 60psi)
Mains power supply
230VAC ±10%, 50/60Hz; 20A fuse
CEE-Plug 230V, 32A
Analyzer dimension
550 x 970 x 600mm
Loader dimension
850 x 350 x 450mm (depending on version)
Analyser weight
approx. 90kg
Loader weight
approx. 55kg
–
The balance should rest on a vibration free support.
Loader dimensions:
Version 36p: 500x250x460mm
Version 130p: 900x500x460
Gas connections:
The supplied tubes carry a connector with G¼" inner diameter ".
Fig. 15: Carrier gas tube
Connections for compressed air;
The tubes supplied together with the analyzer, carry a connector with G¼" inner
diameter.