6.13 Heating elements replacing
DANGER
D2.0005
Mortal danger from electric shock
Exposed power contacts - High Voltage
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
•
Set the mains switch of the analyzer to position 0 and pull the
mains plug of the mains socket.
Fig. 48: Operating procedure for replacement of heating elements
1.
Disconnect the oxygen tube (
1
).
2.
Unscrew the two knurled nuts (
2
) and remove the front assembly (
3
).
3.
Remove the O-ring (
6
) and the safety spring (
7
) from the front end of the ceramic tube.
4.
Pull the dust trap (
10
) with the ceramic combustion tube out of the furnace.
5.
Unscrew and remove the four screws (
4
)
6.
Remove the front stainless panel (
15
)
7.
Unscrew and remove the four screws and (
11
).
8.
Remove the rear stainless cover (
16
).
9.
Remove the clamps (
8
) and (
12
) from the heating elements.
10.
Remove the ceramic spacers (
9
) and (
13
).
11.
Remove the heating elements (
14
).
12.
Re-assemble in reverse order.
NOTICE
Install the clamps (
8
) and (
12
) in a position which allows the heating elements to move at least 5
mm in horizontal (axial) direction.
NOTICE
When you turn on the device after reinstalling, the set setpoint value is driven immediately.
Please check it, see chapter 4.6: Resistance Furnace-Temperature adjustment.
1
3
4
5
2
7
6
4
8
8
9
9
10
11
11
12
13
12
13
14
14
14
16
15
Summary of Contents for CS-2000
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 5: ...5 ...
Page 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Page 80: ...Function description 80 Fig 51 Gas flow system ...
Page 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Page 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...