Maintenance
71
16
Nuts for furnace housing
17
Upper furnace lock
•
Remove the furnace cover by just loosening the nuts (16).
•
Unscrew the knurled nuts (1) and washers (2).
•
Unscrew the wing nuts (3).
•
Unscrew the nuts (4) and detach the tubes (5).
•
Remove the furnace cleaning system, by pulling the mounting (6).
•
Unscrew the nut (7) and the detach tube (8).
•
Unscrew the wing nuts (9) and pull down lower furnace lock (10).
•
Pull-off the lower O-ring (12) from the combustion tube, remove the combustion tube (14)
by pulling it up; remove the upper O-ring (11).
•
Apply a thin layer of grease on the inner surface of the new O-rings (11) and (12), before
mounting them. Apply a thin layer of grease on the outer surface of the combustion tube,
where the new O-rings will be placed.
•
Reinstall in reverse order.