Technical Manual

Temperature- & Humidity Recording System

02.07.96

12:30

Room 1

-23.5°C

Brief Description

MiniMEP Recording Systems (dataloggers) are designed for

paperless

 longtime data recording of temperature values and

(dep. on type) relative humidity. Thus the user is able to get a
proof about the course of temperatures and humidity in
corresponding rooms, 

without

 application of 

additional

components

 as e.g. mechanical printers or PC's.

It is ideal for compliance with 21 CFR process control for food
storage regulations and HACCP Quality Assurance Systems.
The wide temperature range qualifies a MiniMep for both freezing
applications and 'hot' industrial processes.

The Recording System MiniMEP meets all requirements of
the EU Standard EN 12830.

 Functions

z

Records the values of temperature sensors resp. humidity

transmitters for up to 20 years !

Adjustable storage interval.

Example 1: 

781 days with a 5 minute interval

.

Example 2: 

> 6 years with a 15 min. interval

z

Data storage is unlimited even without mains voltage

z

Real time clock with typ. 10 years battery backup

z

Continuous recording by external lead accu possible 

*

z

Alarm too high/low with time delay for each input

z

Monitored sensor inputs

z

Error message memory shows the last 6 occured errors

z

Clear assignment of measuring position, time and value

plain text designation in the display

z

Networkable via interface

z

LCD dot matrix display, backlighted, 4 languages

z

Display shows actual temperature, time and date

z

Lets you view all recorded values any time

z

Easy operating by 3 keys

z

Digital inputs e.g. for door-contact or other messages 

*

z

Isolated alarm relay and built-in buzzer 

*

z

Four display languages selectable:

german, english, french, dutch

 * (not type 404 S, 2404S)

MiniMEP

Nr. 5310965-00/20 E

Please note Operating and

Safety Instructions on page 12 !

Valid for the following types containing Software Version 2.5x:

MiniMEP 404 S, 2404 S
MiniMEP 424 A, 2424 A, 424A/12V
MiniMEP 434 A, 2434 A
MiniMEP 524 A, 2524 A
MiniMEP 534 A, 2534 A

Summary of Contents for MiniMEP 2404S

Page 1: ...minute interval Example 2 6 years with a 15 min interval z Data storage is unlimited even without mains voltage z Real time clock with typ 10 years battery backup z Continuous recording by external le...

Page 2: ...4A 534A canrecord2temperature values and the values of two 4 20 mA humidity transmitters Type Housing Inputs Recording Digital Alarm Limits LC Input for Memory Cap Supply Overview Range inputs Alarm r...

Page 3: ...01 07 96 14 15 sensor 2 21 0 C 01 07 96 14 30 01 07 96 14 45 01 07 96 15 00 P P MINIMEP serial no 000000 01 07 96 14 30 01 07 96 15 00 01 07 96 14 45 correction probe selection push for longer than 1...

Page 4: ...abled during accu operation Real time clock The integrated x tal controlled real time clock can be corrected by only 24 hours Theautomaticsummer winter shiftcanbedisabledforexportpurposes Without main...

Page 5: ...delay is run down an alarm message will be generated 0 01 4 00 h min 1 00 h min name transm 3 individual name for this probe input max 8 characters see table transmitter 3 probe 4 corr 1 the last fou...

Page 6: ...l be generated 0 01 4 00 h min 1 00 h min name probe 3 individual name for this probe input max 8 characters see table probe 3 probe 4 corr 1 the last four corrections of this input with time to date...

Page 7: ...mains ph L report input 2 al2 mains ph alarm relay N O contact Mains neutral alrm relay N C contact alarm relay common PE 1 2 3 4 5 6 7 8 9 RS 232 Interface MiniMEP 4xx RS 232 interface connector The...

Page 8: ...14 input 1 4 20mA 15 input 1 4 20mA 16 probe 2 17 probe 2 com 18 input 3 4 20mA 19 input 3 4 20mA 20 probe 4 21 probe 4 com 22 n c 23 RxD RS 232 24 GND RS 232 25 TxD RS 232 51 R 51 R 11 13 15 17 19 21...

Page 9: ...t A don t need the cables between terminals 37 and 38 TxD data out RxD data in 1 4 2 3 5 Stecker SUB D 9 polig male 7 6 9 8 GND 31 32 16 17 to controller unit PE Mains N Mains Line L Input 1 L Input 2...

Page 10: ...16 17 18 19 ELREHA 63 47 34 35 36 37 38 34 35 36 37 38 93 105 L Eing 1 NK L Eing 2 TK L Ein 3 T r Dimensions of MiniMEP 4xx types Dimensions Terminals of MiniMEP 2 524 A 2 534 A Operating Unit Verbin...

Page 11: ...ler an unlock code must be entered while start up You get this unlock code with the software please see software manual for further information Without this code retrieve and processing of data with t...

Page 12: ...y please contact the manufacturer first Electrical installation and putting into service must be done from authorized personnel Please note the local safety instructions Before installation Check the...

Page 13: ...m relay SPDT isolated 4A cos phi 1 250V Interfaces RS 232 RS 485 Accu interface Chargecontrollerfor1 3AhLeadAccus not types 404 S 2404 S 424 12V Data memory see type overview Real time clock automatic...

Page 14: ...heir companies copy We state the following When operated in accordance with the technical manual the criteria have been met that are outlined in the guidelines of the council for alignment of statutor...

Page 15: ...Drilling template MiniMEP x422 x424 Ma stab 1 1 scale 1 1 116mm 4 57 Bohrer Drill 6mm 24 7mm 27 Bohrer Drill 6mm 24 11 43 163 5 6 42 155 6 1 65mm 2 56...

Reviews: