background image

USER MANUAL

Inteli-Mixer RM-2

6

Using mixing programs

Rotation programs F1, F2, F3, F7
These programs are used to perform mixing of lab-
oratory samples by means of rotation of the rack. 
Rotation speed is indicated on 

RPM display and 

could be changed either during the operation or 
when the rack is stopped.

Tip ! 

Once RPM has been set it will apply to all the pro-

grams in the F menu including customer programs.

Vortexing programs ‘u’ & ‘U’ (Vortexing regimes).
These programs are used to perform intensive mix-
ing of different size test tubes by means of shaking 
and vortexing. Vortexing frequency is indicated on 
RPM display and could be changed either during 
the operation or when the rack is stopped. 
During these regimes rack could be manually posi-
tioned under any angle by hand to insure best per-
formance and convenience.

Tip ! 

Once vortexing frequency has been set it will apply 

to all the programs in the F menu that feature shaking 

including customer programs.
Tip ! 

In case of rack loosing its preset position while 

vortexing reduce the vortexing frequency until opera-

tion becomes stable.

Angular shaking 6, 8, 10, 15, 30, 45, 60, 90, 99
These programs are performing defined angle and 
speed shakes from the vertical initial position. This 
kind of shaking is recommended for mixing sam-
ples in tubes bigger then 5 ml volume.
Current angel settings are displayed on F display. 

Shaking speed is shown on 

RPM display. All the pa-

rameters of angle and speed can be changed either 
during operation or when the rack is stopped. 

Tip !

 In case of rack  loosing its preset position while 

shaking reduce the shaking frequency until operation 

become stable. 

Combined programs F4, F5, F6, F8, F9
These programs are performing different combina-
tions of rotation and vortexing. 
Current speed is shown on 

RPM display and can be 

changed either during operation or when the rack 
is stopped.
Vortexing frequency can be changed by selecting 
on ‘

F’ display either ‘u’ or ‘U’ function and set vor-

texing frequency accordingly.

 (see also Vortexing 

programs ‘u’ and ‘U’)
Now vortexing frequency is adjusted to the best 
performance within current application.

Tip ! 

Vortexing frequency settings that has been de-

fined in ‘u’ or ‘U’ functions and rotation speed settings 

are valid throughout all the programs including cus-

tomer programs.

Customer programs C1, C2, C3
Are used to create customized algorithms of rack 
rotation and shaking 

(see ‘Creating customer pro-

grams’).

Creating customer programs C1, C2, C3

Introduction
Customer programs consist of a sequence of basic 
operations or 

steps* that are inputted by operator 

to the memory of a mixer. Single program can con-
sist of up to 78 steps and must have its start and 
end point in the

 initial point**

*step

 - Is a single basic operation that is inputted in 

the memory by pressing  button.
**Initial point

 - Initial rack vertical position.

List of steps available for creating customer pro-

grams.
SP 

– Rack spinning.

*Clockwise rotation

- Rack rotation clockwise relative-

ly to motor control block.

***Counterclockwise rotation

 - Rack rotation coun-

terclockwise relatively to motor control block
PP

 - Pause Hold the rack in any defined position for 

1.5 sec.
PU

 – Shaking Big amplitude U shaking for 1.5 sec. in 

any defined position.
Pu

 - Shaking Small amplitude u shaking for 1.5 sec. in 

any defined position. 
P0

 – End of program

Tip !

 by inputting any of PP, PU, Pu steps more then 

one time the duration of this operation increases ac-

cordingly.

Summary of Contents for Inteli-Mixer RM-2

Page 1: ...USER MANUAL RM 2L RM 2M RM 2S elminorthamerica com Inteli Mixer RM 2...

Page 2: ...ms F4 F5 F6 F8 F9 6 Customer programs C1 C2 C3 6 Creating customer programs 6 Introduction 6 Entering programming mode 7 Programming mode 7 Learning to use program steps 8 Spinning 8 Temporary pause 8...

Page 3: ...speed rev min from 1 to 99 Speed selection step rev min 1 Environmental temperature C from 2 to 50 Relative moisture of the atmosphere at the temp 20 C not more than 80 Power adapter input 110 220 V...

Page 4: ...on On is on the top of the control block Power supply socket is on the back side of the control block Rack to motor connector is on the right side of the control block Back holder Base platform START...

Page 5: ...ckwise rotation 1 5 sec Hold in initial point F8 110 clockwise rotation 1 5 sec big amplitude U shaking 110 counter clockwise rotation 1 5 sec Hold in initial point 110 counter clockwise rotation 1 5...

Page 6: ...cy until operation become stable Combined programs F4 F5 F6 F8 F9 These programs are performing different combina tions of rotation and vortexing Current speed is shown on RPM display and can be chang...

Page 7: ...s and displays acquire additional functions See figure 2 During programming mode RPM display shows rack rotation angle relatively to initial position See dia gram on fig 4 Program seletion buttons Ste...

Page 8: ...hile entering long programs it is recommended to enter positioning correction each time we pass 00 vertical position point This can be done by selecting SP function in 00 point F display shows P0 and...

Page 9: ...tor that corresponds to 270 degrees see Fig 3 2 9 Press enter button 2 10 Pressing clockwise rotation button or coun ter clockwise rotation button set 00 value on RPM indicator F indicator will show...

Page 10: ...g buttons select operation program see Intelli mixer program overview table Set rotation speed and shaking intensity see using mixing programs Note Operation programs can be selected either during ope...

Page 11: ...emory is empty Enter a new custom program See creating custom programs Disinfection and cleaning Before you start disinfection or cleaning make sure the power adapter is unplugged It is recommended to...

Page 12: ...harges are applied to the con sumer For all further questions concerning exploitation and maintenance please Contact manufacturer or product vendor Certificate of approval Inteli Mixer RM 2 ____ N0 __...

Reviews: