9
USER MANUAL
Inteli-Mixer RM-2
Example 1. Racks rotates clockwise 180 deg. then
counter clockwise 270 deg. Store program in
C1.
1.1 Select
C1 on F display.
1.2 Hold button for 4.5 seconds. Rack will auto-
matically find
00 initial point.
1.3 Pressing clockwise rotation button set 50
value on
RPM indicator, that corresponds to
180 degrees (see Fig. 3).
1.4 Press enter button
1.5 Pressing counter clockwise rotation button
turn the rack until 75 value on
RPM indica-
tor, that corresponds to 270 degrees (see Fig.
3).
1.6 Press enter button
1.7 Pressing clockwise rotation button or coun-
ter clockwise rotation button se
00 value
on
RPM indicator, F indicator will show P0.
1.8 Press enter button
Program is successfully recorded.
Example 2. (Features entering position correction).
Rack rotates clockwise 180 deg., then counter clock-
wise 180 deg., entering position correction in
00
deg. point then rotates 90 deg. counter clockwise.
Store program in
C2.
2.1 Select
C2 on F display.
2.2 Hold button for 4.5 seconds. Rack will auto-
matically find
00 initial point.
2.3 Pressing clockwise rotation button set 50
value on
RPM indicator, that corresponds to
180 degrees (see Fig. 3).
2.4 Press enter button
2.5 Pressing counter clockwise rotation button
turn the rack until
00 value on RPM indica-
tor.
Caution !
reaching 00 point automatically is
offered P0 function that sets end of program. If
you would like to continue programming do not
press button on this step.
2.6 Choose
SP function on F display.
2.7 Press enter button (position correction en-
tered)
2.8 Pressing counter clockwise rotation button
turn the rack until 75 value on
RPM indica-
tor, that corresponds to 270 degrees (see Fig.
3).
2.9 Press enter button
2.10 Pressing clockwise rotation button or coun-
ter clockwise rotation button set
00 value on
RPM indicator, F indicator will show P0.
2.11 Press enter button
Program is successfully recorded.(actions 1.5, 1.6,
1.7 – position correction )
Example 3. Perform small amplitude shaking ‘u’
during 3 sec., then rotate counter clockwise 90 deg.,
then hold 1.5 sec., then shake 1.5 sec. with big am-
plitude ‘
U’ & 1.5 sec. with small amplitude ‘u’, then
rotate clockwise 180 deg., then rotate counter clock-
wise to the
00 deg. point and hold 1.5 sec. Store the
program in
C3.
3.1 Select
C1 on F display.
3.2 Hold button for 4.5 seconds. Rack will au-
tomatically find
00 initial point.
3.3. Select
Pu function on F display.
3.4. Press enter button . 1.5 sec shaking is en-
tered.
3.5. Press enter button . Extra 1.5 sec shaking is
entered.
3.6. Pressing counter clockwise rotation button
turn the rack until 75 value on
RPM indicator,
that corresponds to 90 degrees (see Fig. 3).
3.7. Press enter button .
3.8. Select PP function on F display 1.5 sec. pause.
3.9. Press enter button .
3.10. Select
PU function on F display 1.5 sec. big
amplitude shaking.
3.11. Press enter button .
3.12. Select
Pu function on F display 1.5 sec. small
amplitude shaking.
3.13. Pressing clockwise rotation button turn the
rack until 25 value on
RPM indicator, that cor-
responds to 180 degrees (see Fig. 3).
3.14. Pressing counter clockwise rotation button
turn the rack until
00 value on RPM indicator.
Caution !
reaching 00 point automatically is
offered P0 function that sets end of program. If
you would like to continue programming do not
press button on this step.
3.15. Select
PP function on F display 1.5 sec. pause.
3.16. Press enter button .
3.17. Select
P0 function on F display.
3.18. Press enter button .
Program is successfully recorded.