- 24 -
User’s manual for
CC-411
conductivity meter
v5.30
10.3. Conductivity
measurement
with
manual
temperature
compensation
The measurement with manual temperature compensation may be made in
stable work conditions, e.g. during measurements in laboratory, especially
with use of thermostat, or in case of the temperature probe damage.
Disconnecting the temperature probe switches the meter to manual
temperature compensation.
In case of measurement with manual temperature compensation:
-
connect the conductivity cell to the
F1
connector (pic.2);
-
disconnect the temperature probe;
-
turn the meter on with the
button;
-
choose the conductivity measurement unit according to the subchapter 6.1;
-
if the conductivity cell has not been calibrated earlier, calibrate it according
to the chapter 9;
-
measure the temperature of the solution with thermometer;
-
enter the measured temperature value for manual compensation according
to the point 6.2.1 ;
-
place the cell in the measured solution, it canot touch the walls and the
bottom. The measuring cell should be filled with the sample solution and
shouldn’t include any air bubbles, the electrode’s surface should be evenly
moistened*;
-
after stabilisation read the result (pic.13).
Pic. 13
* - air bubbles may be removed by moving the immersed cell. In order to make moistening of the electrodes
easier, it is recommended to immerse the cell in distilled water with washing-up liquid, and than wash it in
distilled water.
Summary of Contents for CC- 411
Page 1: ...WATERPROOF CONDUCTIVITY METER CC 411 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CC 411 conductivity meter v5 30...
Page 14: ...8 User s manual for CC 411 conductivity meter v5 30 Pic 2...
Page 18: ...12 User s manual for CC 411 conductivity meter v5 30...
Page 19: ...13 II CONDUCTIVITY MEASUREMENT...
Page 20: ...14 User s manual for CC 411 conductivity meter v5 30...
Page 34: ...28 User s manual for CC 411 conductivity meter v5 30...
Page 35: ...29 III TEMPERATURE MEASUREMENT...
Page 36: ...30 User s manual for CC 411 conductivity meter v5 30...
Page 38: ...32 User s manual for CC 411 conductivity meter v5 30...
Page 39: ...33 IV OTHER...
Page 40: ...34 User s manual for CC 411 conductivity meter v5 30...
Page 45: ...39...
Page 46: ...40 User s manual for CC 411 conductivity meter v5 30...
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......