8. CHOICE AND MAINTENANCE OF THE CONDUCTIVITY CELL
8.1. Choice of the conductivity cell
Conductivity
in
CC-411
meter
can
be
measured
in
range
0 – 100 mS/cm. The meter co-operates with conductivity cells with K constant
= 0.050 ÷ 1.500 cm-1 and equipped with BNC-50 connector. Depending on
the required measuring range it is necessary to choose appropriate cell with K
constant which enables obtaining accurate results. Beyond the range the cell
looses its linearity and the results have greater error. Accurate measurements
in the whole range are possible with use of 2 different conductivity cells.
In range 1 – 100 mS/cm the standard conductivity cell with K constant =1 cm-
1 should be used.
For measurements of ultra pure water with conductivity below 1
µ
S/cm a cell
with K constant = 0.1 cm-1 should be used.
Depending on the expected measuring range, choose the appropriate cell
using the chart below (Pic.6).
Pic. 6. Dependence between the measuring range and of applied conductivity
cells.
Summary of Contents for CC- 411
Page 1: ...WATERPROOF CONDUCTIVITY METER CC 411 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CC 411 conductivity meter v5 30...
Page 14: ...8 User s manual for CC 411 conductivity meter v5 30 Pic 2...
Page 18: ...12 User s manual for CC 411 conductivity meter v5 30...
Page 19: ...13 II CONDUCTIVITY MEASUREMENT...
Page 20: ...14 User s manual for CC 411 conductivity meter v5 30...
Page 34: ...28 User s manual for CC 411 conductivity meter v5 30...
Page 35: ...29 III TEMPERATURE MEASUREMENT...
Page 36: ...30 User s manual for CC 411 conductivity meter v5 30...
Page 38: ...32 User s manual for CC 411 conductivity meter v5 30...
Page 39: ...33 IV OTHER...
Page 40: ...34 User s manual for CC 411 conductivity meter v5 30...
Page 45: ...39...
Page 46: ...40 User s manual for CC 411 conductivity meter v5 30...
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......