Operation manual ISOTEST 4 S / 4 S
p l u s
Page 17
alternatively
Required accessories:
Grounding cable Y - type
(clamping tongs - clamping tongs/plug)
Part number 0174250010
Application:
1. Connect the plug of the Y grounding cable to the grounding
receptacle of the ISOTEST
®
test equipment.
2.
Connect the clamping tongs at the other end of the grounding cable
to the test object.
Connect the second pair of clamping tongs to a ground terminal.
3. Turn the ISOTEST
®
test equipment on and perform the high-voltage
test.
alternatively
Required accessories:
Standard grounding cable with plug and clamping tongs
Part number: 0174320000
(supplied)
Grounding stick with cable, 2 m
Part number 0174200000
Application:
1. Connect the plug of the standard grounding cable to the grounding
receptacle of the ISOTEST
®
test equipment.
Connect the clamping tongs at the other end of the grounding cable
to the test object.
2. Connect the clamping tongs at the cable from the grounding stick to
an accessible area of the test object.
3.
Deeply bury the grounding stick in the soil. If the soil is dry,
thoroughly water the respective area to obtain a low-resistance
connection.
4.
Turn the ISOTEST
®
test equipment on and perform the high-voltage
test.