operating it, afterwards.
Due to over discharging, the battery pack
Repeat the charging and discharging
procedures for several times. If the result is
damaged.
the same, change the battery pack.
The battery pack has completed working life.
Repeat the charging and discharging
procedures for several times. If the result is
the same, change the battery pack.
8.1 Battery
LIFETIME SA-PB device is equipped with high quality nickel metal hydrate (Ni-MH) batteries.
8.1. Changing battery group
Important!
In order to avoid any possible damages on electronic circuit, do not touch the electronic board.
1.
Shut LIFETIME SA-PB device down.
2.
Remove the device from external supply voltage.
3.
Remove the container and other accessories.
4.
Unscrew 4 star-head screws on the battery chamber
cover on the back of the body.
5.
Carefully disconnect the battery pack socket
connector.
6.
Replace the flat battery pack.
7.
Wait about half a minute before installing the new
battery pack. Thus, capacitors on the printed circuit board
are discharged.
8.
Place the new battery pack into its slot.
9.
Carefully insert the battery pack socket
connector.
10.
Close the battery slot cover and tighten the screws.
11.
Calibrate the device.
Unless the electronic control circuit is in sync with the
battery pack, the red LED of the capacity indicator flashes.
The LIFETIME SA-PB device is ready to run when the
battery pack is full, but the charge status of the battery pack
will not be displayed unless calibration is performed..
12.
Do a functional check (see the chapter
"7.2 Function Controlling")
8.1.2 Calibrating capacity indicator
With the calibration procedure, the capacity indicator and energy amount of the battery pack become equal with each other.
The following condition require for the calibration:
After each battery pack replacament,
After each fuses changed,
Please follow those steps for calibration:
1.
Discharge LIFETIME SA-PB device until it automatically stops.
2. For charging, connect LIFETIME SA-PB device to the external power. The charging procedure lasts for approximately 2.5 hours. During
discharging, d
ue to the high residual capacity, it is possible for the battery pack to heat up if the pump runs for a long time. In this case,
charging starts when the temperature of the battery pack drops below 40 ° C. Depending on the capacity, this process can take up to 45
minutes.
9. Maintenance
9.1 Periods
LIFETIME SA-PB device does not need any maintenance. However, please consider the periods for the regular functions ("See the
section 7.1 Periods"). We recommend that you calibrate the battery pack every 8 weeks in accordance with section 8.1.2 to maintain the
battery pack operating mode and the lifespan of the battery pack. In this context, the battery group is discharged in a controlled manner and
then recharged. We recommend that you carry out maintenance work, such as inspection and repair, to the manufacturer company
ELMASLAR or a qualified specialist.
9.2 Disposal
Page
12
/
14