4.2 Assembling a Bag 

  

If the accessory bag is not included in the scope of delivery of your device, you can order the accessory bag as an accessory. This bag is 

used to protect aspiration catheters and other small parts. 

 

1- Mount the accessory bag on the side of the motor unit using velcro straps. 

 

5.

 

Usage 

 
 

5.1 First Operating 

  

Recommendation: 

When you use the reusable collecting container, it will 

be very convenient for you to fill the container with approximately 50-
100 ml of disinfection liquid or water before starting the aspiration 
application. 

The vacuumed particles will not easily stick to the bottom of the container 

thanks to this method

 

1.

 

Remove the aspiration tube from the hose wrap plate. 

 

2.  If necessary, adapt an aspirating catheter of the size suitable for 

tracheal or nasopharyngeal aspiration to the end piece.. 

 

       Important! 

 

To effectively aspirate viscous and solid food particles from the oral cavity, remove the 

piece attached to the ends of the aspiration tube and use only the aspiration tube 

 

3.

 

Press the 

O/I

 button. After the device is opened, it starts in the lowest vacuum 

level. 

 

4.

 

Select the vacuum power by turning the switch button. 

 

5.

 

LIFETIME SA-PB is ready for operation now and you can start the aspiration 

procedure. 

 

5.2 Aspiration 

 

During the aspiration process, be careful not to cause any injury to the patient's mouth and pharynx, especially mucous membranes. 

 

 When the device is fully charged it can operate in the highest level of vacuuming for minutes continuously. By shutting the device 

down after a 45 minute of continual operation, you can avoid the overheating of the device. Please wait for at least 2 hours for cooling the 

device. 

Important! 

 

When operating with the multi-use collecting container, the device must always stand upright, because otherwise the ball of the overfill 

assembly cannot safely close the connection to the suction pump. As a result, secretions can enter the suction pump and damage the 

pump. 

 

If operated with a container used once, in order for avoiding the obstruction of the overflow valve integrated into suction bag by the 

saliva, the device should always be standing vertically. If the fluid is scattered onto the overflow valve, it becomes air resistant and the 

disposable aspiration bag should be changed. 

 

Information

! By monitoring the collecting container throughout the suction process, 

Please check. Empty the collecting container at 950 ml level so that the overfill assembly does not get dirty and therefore does not need 

to be cleaned. 

If the overfill prevention device is activated before the filling limit is reached during aspiration application, you should take a short break 

from the aspiration process and bleed the system. 

 

Information

! Please also regularly check the charging status of the battery pack during aspiration application. As soon as the red warning 

LED lights up in the capacity indicator, aspiration application should be continued by using a backup device. If the battery pack continues to 

discharge, the battery pack may be damaged by excessive discharge and a significant reduction in capacity may occur. 

 

Discharging air from the system 

1.

 

The connection between the device and the collector is removed from the collector side. 

 

2.

 

Wait until the float of the overfill prevention device falls back. 

3.

 

The connection between the device and the container is reconnected. You can continue the aspiration procedure. 

Emptying the reusable collection container 

 

When the reusable collection container is filled to the fullness  limit, you should stop the aspiration process and empty the  collection 
container. 
 

Important! 

When  removing  and  emptying  the  collection  container,  please  be  careful  not  to  accidentally  come  out  of  the  collection 

container and not to spill the contents.

 

1.

 

Loosen the aspiration tube from the hose wrap plate in order to remove the container more easily and prevent the secretion cap from 

accidentally detaching. 

 

2.

 

Separate the container from the motor unit. 

3.

 

Carefully pull out the secretion cap. 

4.

 

Empty the container. 

Important!

 For disposal please observe the relevant instructions. 

5.

 

Insert the secretion cap on the container again. 

        

6. 

Hang the collector back onto the motor unit. Please make sure that the colelctor is pressed onto the motor unit and snaps into the 

locking slot.

 

Page 

8

 / 

14 

Summary of Contents for LIFETIME SA-PB 1

Page 1: ...LIFETIME SA PB Suction Device Product Descriptions and Operating Manual ELMASLAR A MODELS LIFETIME SA PB 1 LIFETIME SA PB 2 LIFETIME SA PB 3 LIFETIME SA PB 4 Page 1 14...

Page 2: ...n 8 5 3 Ending Aspiration 9 5 4 Charging LIFETIME SA PB 9 6 Hygienic Preparation 9 6 1 Preparation 9 6 2 Cleansing Disinfection and Sterilization 9 6 3 Connecting Pieces 10 7 Controlling Functions 10...

Page 3: ...1 Overview Front view of LIFETIME SA PB device with reusable collecting container Back view of LIFETIME SA PB device Page 3 14...

Page 4: ...Reusable Collection Container Accessories and spare parts Page 4 14...

Page 5: ...visual warning for 6 seconds 2 Product Descriptions 2 1 Intended Use LIFETIME SA PB is a mobile portable electrically operated medical suction pump used temporarily in adults children and infants To...

Page 6: ...ing with the multi use collection container the device must always stand upright because the ball of the overfill assembly cannot safely close the connection to the suction pump and secretions may ent...

Page 7: ...template on the wall and align it with a water gage 3 Mark the holes for the assembly points 4 Make the holes by using a 3 mm sized drill bit 5 Screw the wall stabilizer by using the screws in the pac...

Page 8: ...ve integrated into suction bag by the saliva the device should always be standing vertically If the fluid is scattered onto the overflow valve it becomes air resistant and the disposable aspiration ba...

Page 9: ...cess ends all the lights on the indicator band will light up steadily Charging with wall stabilizer 1 Connect LIFETIME SA PB device to the wall stabilizer 2 Charging starts automatically a few seconds...

Page 10: ...d is 20 minutes 6 3 Connecting Pieces Following the cleaning disinfection or sterilization procedures connect the pieces as follows 1 Screw the float of the overfill prevention device into the secreti...

Page 11: ...are Correct the incorrect polar connections and reverse change the fuses if required Connection between circle plate and battery Do the interconnection entirely group is not well inserted Device is no...

Page 12: ...ck is full but the charge status of the battery pack will not be displayed unless calibration is performed 12 Do a functional check see the chapter 7 2 Function Controlling 8 1 2 Calibrating capacity...

Page 13: ...vacuum pressure in 12 V 650mmHg in sea level Aspiration hose 7 12 mm Length 1500 mm reusable container Bacteria filter Filter component of 1 m Motor power 55 W Rated voltage 12 V Maximum power 8 A Whi...

Page 14: ...ransport damage caused by improper packaging when the device is returned Not performing the required or stipulated maintenance procedures Normal erosion due to operating The following parts are under...

Reviews: