Operating faults
92
BA_Elmasolvex_VA_EN_Vers.05.2018
© Elma Schmidbauer GmbH
•
Suction filter in the media tank (
Fig.
3.13.1.G
) clogged / blocked?
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
217
Abortion
Error!
Cleaning chamber
drains too slowly.
Wait, then remove and
clean sieve insert.
Timeout 2 of the Empty
level
fill level sensor during
draining.
Check the following possibilities:
•
Condensate separator or activated
carbon filter (if connected) clogged?
•
Exhaust air tube (
chapter 4.3
) kinked?
•
Air intake and exhaust openings in the
cleaning chamber (
Fig. 3.12.1.A
)
clogged?
•
Sieve insert in the cleaning chamber
clogged (
Fig.9.1.4.1
)?
•
Can the valves of the quick couplings
(
Fig. 4.4.1.2.J
) be pressed in / moved
correctly?
•
Can the valves of the quick couplings
on the media tanks (
Fig. 4.4.1.1.F
) be
pressed in / moved correctly?
•
Are the quick couplings on the media
tanks (
Fig. 3.13.1. B/C
) correctly
engaged?
•
Suction filter in the media tank (
Fig.
3.13.1.G
) clogged / blocked?
•
Filling level sensor
Empty level
(installed under the sieve insert)
(
Fig.9.1.4.1.F
) soiled? Clean with
suitable cloth if necessary.
•
Fill level sensor
Empty level
(
Fig.9.1.4.1.F
) damaged (scratches,
cracks)?
•
Retrofit ant-reflection sleeve for
Empty
level
fill level sensor.
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
218
Abortion
Error!
Cleaning chamber does
not drain.
Timeout of the working level
fill level sensor during
draining of the cleaning
chamber.
Check the following possibilities:
•
Condensate separator, activated
carbon filter (if connected) clogged?
•
Exhaust air tube (
chapter 4.3
) kinked?
•
Air intake and exhaust openings in the
cleaning chamber (
Fig. 3.12.1.A
)
clogged?
•
Sieve insert in the cleaning chamber
clogged (
Fig.9.1.4.1
)?
•
Fill level in the media tank overfilled?
•
Can the valves of the quick couplings
(
Fig. 4.4.1.2.J
) be pressed in / moved
correctly?