Before the initial commissioning
26
BA_Elmasolvex_VA_EN_Vers.05.2018
© Elma Schmidbauer GmbH
Only permitted operating materials must be used for safety
reasons and to prevent cleaning machine damage.
The media tanks must only be removed when the cleaning
machine is switched off (empty cleaning chamber) for safety
reasons.
Observe the instructions for recommended operating materials
and restrictions for operating materials (
Chapter 8
).
Fire and explosion hazard!
When handling flammable materials, observe the applicable
safety regulations according to the safety data sheet of the
respective solvent.
Keep all kinds of ignition sources away.
Prevent ignition sparks from electrostatic discharge. Discharge
possible electrostatic charges (body charge) by touching any
grounded equipment before handling flammable materials: e.g.
water tap, metal surface of the case of the cleaning machine or
use ESD protective equipment (ESD arm band).
If liquid escapes while replacing the media tank, this must only
be removed using a dry cloth (to prevent electrostatic
charging).
Start with the media tank #1 (
Fig. 4.4.1.#1
):
1. Push up the covers for the media tanks (
Fig. 4.4.2.A
/
A
).
2. Unlock the shut-off valves for the tube connections:
Press the respective unlocking button on the shut-off valve
of the media tank (
Fig. 4.4.3.B
/
C
) and pull the tube
connection upwards out of the shut-off valve at the same
time.
3. Remove the media tank from the cleaning machine using
the recessed handle.
4. Open the screw cap and fill the media tank with the suitable
cleaning medium up to a filling level between the
Min
–
Max
markings (
Fig. 4.4.3.D
).
Only use permitted
operating materials.
Do not remove
during operation.
DANGER
Procedure for filling