Important safety instructions
© Elma Schmidbauer GmbH
BA_Elmasolvex_SE_EN_Vers. 03.2018
5
Information sticker on the cleaning machine with brief
instructions to be observed:
Only operate with all 4 media tanks
Check fill levels of the media tanks
Check top case fan daily before start cleaning
Switch off cleaning machine / fan at the main switch (e.g. after
finishing work, Emergency Stop)
Heater can be interrupted at the
STEP TIME
rotary knob
2.1.2
Signal words in this manual
The "danger" signal word warns about severe injuries with risk
of fatal injury.
The "warning" signal word warns about severe injuries.
The "caution" signal word warns about slight to medium injuries.
The "attention" signal word warns about property damage.
2.2
Safety instructions for use of the machine
First, strictly make yourself familiar with the safety instructions
before start-up.
A summary of the safety instructions can be found here. These
are shown again in this operating manual before the respective
handling instructions.
The machine must only be operated by trained personnel in
accordance with this operating manual.
This watch cleaning machine is exclusively intended for
cleaning mechanical precision parts such as dismantled and not
dismantled watch movements using solvent and aqueous
solution based cleaning and rinsing media (
see Chapter 7
).
The flammable media used must also satisfy the conditions of
flashpoint >= 23 °C and ignition temperature >= 200 °C.
Flammable liquids can be used as cleaning and rinsing media
under compliance with the indicated conditions.
In the case of intended use of flammable solvent-based media,
4 air changes
1)
per hour and per cleaning machine using
technical ventilation measures must be guaranteed for the
explosion-safe operation of the cleaning machine(s).
1)
must be realised by correctly designed technical ventilation installed
on the building side (e.g. suitable exhaust air blower into the open air)
It is not permitted to operate the machine in an environment
with potentially explosive atmosphere of flammable gases,
vapours or dusts
The permitted ambient temperature is +5 °C to 30 °C.
Danger
Warning
Caution
Attention
User
Intended use
Air change in
installation area
Ambient conditions
Ambient temperature