Wire Map measure procedure.
1. Select cable type under test.
2. Connect the cable under test to the appropriate connector on
the tester.
3. Press WIRE MAP key.
Good cable with remote unit
Wire Map Failure Using remote unit (ID).
FAILURE DISPLAY
WIRING
DESCRIPTION
Short
(near
end)
1
3
6
2
1
3
6
2
Alternately display
an “s” with actual
wire number of
each wire shorted.
Open
2
6
3
2
6
3
Alternately display
“o” with the
number of each
open wire.
Miswire
3
3
2
2
Displays the wiring
detected by the
test tool and flash
the wire numbers
involved in the
anomaly.
VII.
DATA MEMORY AND READ OPERATING
1. Press MR key each time, stores one set logged data in
memory, LCD shows M and memory location numbers (01 to
99). If no test data can not perform this mode.
2. Press MR key for 3 seconds to enter the read memory data
mode, LCD shows R and memory location numbers, additional
information can be viewed with the ▲/ ▼key.
3. Press
key to scroll through the logged readings.
4. Press MR key again to exit READ mode, LCD will show “Out r
EAD”.
5. Erase the memory of dataloggeer.
Press
key to turn off the meter.
Press and hold down MR key then press
key to turn on the
meter, until LCD shows “dEL”, press ▲▼ key to cycle display “YES”
or “NO”, select “YES” and press
key to cleared all memory.
VIII.
MAINTENANCE
1. Cleaning :
Periodically wipe the case with a damp cloth and mild detergent.
Do not use abrasives or solvents. Clean and dry as required.
2. Battery replacement :
When LCD display shows “
+
”, the power of
battery has already been in-sufficient to support an
accurate test. At this moment replace it with new
battery from the battery compartment.