Maintenance 

  Caution 

To avoid electric shock, personal injury, or damage to the Calibrator: 

l

 

Do  not  service  this  product  other  than  as  described  in  this 

Instruction  Sheet  unless  you  are  a  qualified  technician  and  have 

the required equipment and service information. 

l

 

Remove any input signals prior to removing test leads and opening 

case.

 

l

 

When  servicing  the  Calibrator,  use  only  specified  replacement 

parts.

 

l

 

Do not allow water to get in the case.

 

 

In Case of Difficulty 

l

 Make  sure  you  are  using  the  Calibrator  as  described  in  this  Instruction 

Sheet. 

l

 Check the battery and test leads. Replace as needed. 

If  the  Calibrator  is  under  warranty,  refer  to  the  warranty  statement  for  warranty 

terms, conditions, and product return information. 

If the warranty has lapsed, the Calibrator will be repaired and returned for a fixed 

fee. 

 

Cleaning 

Periodically  wipe  the  case  with  a  damp  cloth  and  detergent;  do  not  use 

abrasives or solvents. 

 

Calibration 

Calibrate  the  Calibrator  once  a  year  to  ensure  that  it  performs  according  to  its 

specifications. 

12

 

 

Measuring dc mA with LOOP POWER 

  Caution

 

To prevent damage to the unit under test, ensure that the   

Calibrator is in the correct mode before connecting the test leads. 

Loop  power  provide  +  24V  to  power  a  transmitter  and  to  read  loop  current 

simultaneously. 

To measure dc mA with LOOP POWER: 

1. Press                    to step to 

LOOP POWER

 mode. 

2. 

MEASURE mA

 and 

LOOP POWER

 are displayed. 

3. Touch test lead probes to the circuit across the load or power source as 

shown below. 

To exit 

LOOP POWER

, change measurement mode. 

 

Summary of Contents for 30C

Page 1: ...Loop Calibrator Instruction Sheet ...

Page 2: ...s that can lead to electric shock l Remove test leads from the Calibrator before opening the case or battery door Symbol Symbol Meaning Earth ground Caution Important information Refer to instruction sheet Double insulated Battery Conforms to European union requirements Direct current 3 ...

Page 3: ...to 24mA and 4 to 20mA current loops and instruments In SOURCE mode the Calibrator supplies the current In SIMULATE mode the Calibrator simulates a 2 wire transmitter in an externally powered current loop Changing the mA Output Span The Calibrator has two mA output spans l 4mA to 20mA 0 to 100 default l 0mA to 24mA 25 to 125 optional Sourcing mA Use SOURCE mode to supply current to a passive circui...

Page 4: ...itter at the same time with carry a DMM l Innovation output adjustment dial with 1mA and 100mA resolution l Storage temperature 40 C to 60 C l Operating temperature 10 C to 55 C l Power requirements Single 9V battery ANSI NEDA 1604A or IEC 6LR61 l Battery life typical Approx 15 hours l Size 88mm W 178mm L 33mm H 3 5 in W 7 in L 1 3 in H l Weight 300g with holster 14 Simulating a Transmitter When s...

Page 5: ...atement for warranty terms conditions and product return information If the warranty has lapsed the Calibrator will be repaired and returned for a fixed fee Cleaning Periodically wipe the case with a damp cloth and detergent do not use abrasives or solvents Calibration Calibrate the Calibrator once a year to ensure that it performs according to its specifications 12 Measuring dc mA with LOOP POWER...

Page 6: ...e Calibrator before use Do not use it if appears damaged l Check the test leads for continuity damaged insulation or exposed metal Replace damaged test leads l Never apply more than 40V between any two terminals or between any terminal and earth ground l Use the proper terminals mode and range for your measuring or sourcing application l To prevent damage to the unit under test put the Calibrator ...

Page 7: ...rrent in 1mA steps simply turn the knob l To adjust the current in 200mA steps push in and turn the knob Press to step the current up 25 of full scale 20mA At full scale press to step the current down 25 of full scale Press simultaneously to enter the Auto Ramp mode and select a ramp form A continuously supplied or controlled mA ramping signal is produced in one of three ramp forms slow fast or st...

Page 8: ...0 05 rdg 5dgts up to 15V non specification 20V 1 TX SIMULATOR 2wire transmitter simulation External excitation voltage 12V min 30V max The current sink levels are adjustable with accuracies as in the source OUTPUT RAMP Current Ramp 4 to 20 on all ranges Ramp times 50sec 20sec 5sec 3 ramp type Ramp operation is also available in TX SIMULATOR function 15 ...

Page 9: ...0 to 100 to 0 stair ramp in 25 steps pausing 5 sec at each step 8 Replacing the Battery Warning To avoid false readings which could lead to electric shock or injury replace the battery as soon as low battery indicator appears on the display Use only a single 9V battery properly installed to power the Calibrator The Calibrator uses a single 9V alkaline battery ANSI NEDA 1604A or IEC 6LR61 To replac...

Page 10: ...MEASURE mA is displayed 2 Touch test lead probes to the circuit across the load or power source as shown below 10 Measuring dc Volts Caution To prevent damage to the unit under test ensure that the Calibrator is in the correct mode before connecting the test leads To measure dc Volts 1 Press to step to MEASURE mode MEASURE V is displayed 2 Touch test lead probes across the load or power source 11 ...

Reviews: