MEASURE / DC VOLTAGE(auto) 

Range 

Resolution 

Accuracy 

Overload Protection 

2V 

100

 

±

(0.05% rdg + 5dgts) 

100V DC or AC peak with 

10 Seconds 

20V 

1

 

50V 

10

 

 

MEASURE / DC CURRENT(auto) 

Range 

Resolution 

Accuracy 

Overload Protection 

2

 

0.1

 

±

(0.05% rdg + 5dgts) 

unfused 

20

 

1

 

25

 

10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16 

 

Using the mA Sourcing (Output) Modes 

The  Calibrator  outputs  current  for  calibrating  and  testing  0  to  24mA  and  4  to 

20mA current loops and instruments. 

In 

SOURCE

 mode, the Calibrator supplies the current. 

In  SIMULATE  mode,  the  Calibrator  simulates  a  2-wire  transmitter  in  an 

externally-powered current loop.

 

 

Changing the mA Output Span 

The Calibrator has two mA output spans: 

l

 4mA to 20mA ( 0% to 100% ) [default] 

l

 0mA to 24mA ( -25% to 125% ) [optional] 

 

Sourcing mA 

Use 

SOURCE

 mode to supply current to a passive circuit. 

A  path  must  exist  for  current  to  flow  between  the  +  and 

COM

  terminals. 

Otherwise the display indicates an

 

Display 

Flicker

 when you set an output value.

 

 

 

 

 

 

 

 

 

 

Summary of Contents for 30C

Page 1: ...Loop Calibrator Instruction Sheet ...

Page 2: ...s that can lead to electric shock l Remove test leads from the Calibrator before opening the case or battery door Symbol Symbol Meaning Earth ground Caution Important information Refer to instruction sheet Double insulated Battery Conforms to European union requirements Direct current 3 ...

Page 3: ...to 24mA and 4 to 20mA current loops and instruments In SOURCE mode the Calibrator supplies the current In SIMULATE mode the Calibrator simulates a 2 wire transmitter in an externally powered current loop Changing the mA Output Span The Calibrator has two mA output spans l 4mA to 20mA 0 to 100 default l 0mA to 24mA 25 to 125 optional Sourcing mA Use SOURCE mode to supply current to a passive circui...

Page 4: ...itter at the same time with carry a DMM l Innovation output adjustment dial with 1mA and 100mA resolution l Storage temperature 40 C to 60 C l Operating temperature 10 C to 55 C l Power requirements Single 9V battery ANSI NEDA 1604A or IEC 6LR61 l Battery life typical Approx 15 hours l Size 88mm W 178mm L 33mm H 3 5 in W 7 in L 1 3 in H l Weight 300g with holster 14 Simulating a Transmitter When s...

Page 5: ...atement for warranty terms conditions and product return information If the warranty has lapsed the Calibrator will be repaired and returned for a fixed fee Cleaning Periodically wipe the case with a damp cloth and detergent do not use abrasives or solvents Calibration Calibrate the Calibrator once a year to ensure that it performs according to its specifications 12 Measuring dc mA with LOOP POWER...

Page 6: ...e Calibrator before use Do not use it if appears damaged l Check the test leads for continuity damaged insulation or exposed metal Replace damaged test leads l Never apply more than 40V between any two terminals or between any terminal and earth ground l Use the proper terminals mode and range for your measuring or sourcing application l To prevent damage to the unit under test put the Calibrator ...

Page 7: ...rrent in 1mA steps simply turn the knob l To adjust the current in 200mA steps push in and turn the knob Press to step the current up 25 of full scale 20mA At full scale press to step the current down 25 of full scale Press simultaneously to enter the Auto Ramp mode and select a ramp form A continuously supplied or controlled mA ramping signal is produced in one of three ramp forms slow fast or st...

Page 8: ...0 05 rdg 5dgts up to 15V non specification 20V 1 TX SIMULATOR 2wire transmitter simulation External excitation voltage 12V min 30V max The current sink levels are adjustable with accuracies as in the source OUTPUT RAMP Current Ramp 4 to 20 on all ranges Ramp times 50sec 20sec 5sec 3 ramp type Ramp operation is also available in TX SIMULATOR function 15 ...

Page 9: ...0 to 100 to 0 stair ramp in 25 steps pausing 5 sec at each step 8 Replacing the Battery Warning To avoid false readings which could lead to electric shock or injury replace the battery as soon as low battery indicator appears on the display Use only a single 9V battery properly installed to power the Calibrator The Calibrator uses a single 9V alkaline battery ANSI NEDA 1604A or IEC 6LR61 To replac...

Page 10: ...MEASURE mA is displayed 2 Touch test lead probes to the circuit across the load or power source as shown below 10 Measuring dc Volts Caution To prevent damage to the unit under test ensure that the Calibrator is in the correct mode before connecting the test leads To measure dc Volts 1 Press to step to MEASURE mode MEASURE V is displayed 2 Touch test lead probes across the load or power source 11 ...