MEASURE / DC VOLTAGE(auto)
Range
Resolution
Accuracy
Overload Protection
2V
100
㎶
±
(0.05% rdg + 5dgts)
100V DC or AC peak with
10 Seconds
20V
1
㎷
50V
10
㎷
MEASURE / DC CURRENT(auto)
Range
Resolution
Accuracy
Overload Protection
2
㎃
0.1
㎂
±
(0.05% rdg + 5dgts)
unfused
20
㎃
1
㎂
25
㎃
10
㎂
16
Using the mA Sourcing (Output) Modes
The Calibrator outputs current for calibrating and testing 0 to 24mA and 4 to
20mA current loops and instruments.
In
SOURCE
mode, the Calibrator supplies the current.
In SIMULATE mode, the Calibrator simulates a 2-wire transmitter in an
externally-powered current loop.
Changing the mA Output Span
The Calibrator has two mA output spans:
l
4mA to 20mA ( 0% to 100% ) [default]
l
0mA to 24mA ( -25% to 125% ) [optional]
Sourcing mA
Use
SOURCE
mode to supply current to a passive circuit.
A path must exist for current to flow between the + and
COM
terminals.
Otherwise the display indicates an
Display
Flicker
when you set an output value.
5