Magnetic length and angle measuring system
MPI-15
ELESA models all rights reserved in accordance with the law. Always mention the source when reproducing our drawings.
ITA
ENG
FRA
DEU
Parameter
Description
Available options
Std Value
dir
Counting Direction
Press to choose between
-dir
(negative) and
dir-
(positive).
-dir
OriGin
Reference Value
The parameter value is shown on the display. Press to increase the flashing number/
mark. Press to select the next number. Programmable value
0.00000 +/- 9.99999
.
0.0
OFFSEt
Offset Value
The parameter value is shown on the display. Press to increase the flashing number/
mark. Press to select the next number. Programmable value
0.00000 +/- 9.99999
.
0.0
UnitS
Unit of measure
The parameter enables or disables the units of measure key in the operating mode.
Press or to choose between:
EnAbLE
: the unit of measure can be changed by pressing the key
in the operating mode.
diSAbLE
: the unit of measure cannot be changed in the operating mode.
diSAbLE
rES
Resolution
The parameter allows defining the resolution of the display.
Press or to choose between the available options:
mm (0.1), inch (0.001),
degrees (0.01), degrees (0.1), degrees (1)
.
mm 0.1
Lin_cor
Linear scale
coefficient
The parameter value is shown on the display. Press to increase the flashing number/
mark. Press to select the next number. Programmable value
0.00001 +/- 9.99999.
0.00000 cannot be accepted (the coefficient is automatically set to 1.00000).
1.00000
AnG_cor
Angular scale
coefficient
The parameter value is shown on the display. Press to increase the flashing number/
mark. Press to select the next number. Programmable value
0.00000 +/- 9.99999.
0.00000 cannot be accepted (the coefficient is automatically set to 1.00000).
See Par. 8 for angle measurement description.
1.00000
AutoOFF
Automatic Switch-off
The parameter enables or disables the automatic switch-off.
Use and to choose between:
EnAbLE
: the display automatically switches off after the selected period
of time (parameter
t-out
).
diSAbLE
: display does not switch-off automatically.
diSAbLE
t-out
Switch-off time
Use and to choose between the available values (in hours).
Select
0.2;0.5;1;2;4;8
(in hours). When unused, the display automatically switch off after
the
switch-off time
.
0.2 h
on-OFF
Switch-off mode
The parameter allows defining the switch-off mode.
Use and to choose between:
dirECt
: the display switches off immediately by pressing
dEL’d
: the display switches off by pressing for about 4 seconds
dirECt
LoadOrG
Reference value
loading mode
The parameter allows defining the reference value loading mode.
Use and to choose between:
EnAbLE
: in the absolute mode (
ABS
) the absolute reference value is loaded
by pressing
diSAbLE
: in the absolute mode (
ABS
) the absolute reference value is loaded
by pressing +
diSAbLE
CLr_reL
Incremental counter
reset mode.
The parameter allows defining the incremental counter reset mode.
Use and to choose between:
EnAbLE
: in the incremental mode (
◀◀◀
) the incremental counter is set to zero
by pressing
diSAbLE
: in the incremental mode (
◀◀◀
) the incremental counter cannot be set to zero
(the incremental counter is set to zero when the incremental mode is selected).
diSAbLE