Data
E-15 ElektroPhysik
〖
6.4.9
〗
Gauge data
In this function, users can inquiry the instrument's data:
〖
6.4.10
〗
Probe calibration
For the measuring error caused by the probe replacement and other factors
and it can not eliminated by using the “Auto calibration” function. (Please
make sure that the error is not caused by the incorrect sound velocity
setting), at that moment by the function of “second calibration”, the error
will eliminated according to the system error. The steps are as follows:
1.
Adjust the sound velocity at 5920 m/s
2.
Measure the thickness of the standard delivered test block. The
measured value should be 4.00mm ±0.01mm. If the measured value is not
that, the usage of “Probe calibration” can eliminate the error.
3.
Adjust the measured value and make it meets the above requirements.
〖
6.4.11
〗
Total reset
During the usage, when users can not ensure why the problems comes out
and with some questions on setting, this function can be used to make the
parameters to reset the factory status to eliminate any abnormal because of
the parameters setting.