Page 8 of 20 M6627 Issue 7.0
5.
INSTALLATION.
5.1. Electrical safety and installation.
5.1.1. This equipment is designed to safe under the following conditions:-
Indoor use.
Altitude up to 2000 meters.
Temperatures between 5
C and 40
C.
Maximum relative humidity 80% for temperatures up to 31
C decreasing linearly to 50%
relative humidity at 40
C.
Mains supply voltage fluctuations up to
10% of the nominal voltage.
Transient over voltages typically present on the mains supply.
Applicable rated pollution degree 2.
5.1.2.
This equipment must be earthed / grounded to a fixed earth / grounded mains
socket outlet. The mains supply is to be earthed / grounded in accordance with current
legislation.
5.1.3. Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure the
correct mains cable fuse if fitted). See Technical Information Section 8 of the Instruction
book.
5.1.4. Check the voltage on the product data label of this unit. Ensure the rating conforms to
your local supply.
5.1.5. It is recommended this unit be connected to a mains supply source which incorporates
an RCD or GFCI device.
5.1.6. The unit is supplied with a moulded plug and mains lead set wired as follows.
Green / Yellow
or
Green
=
Earth / Ground
Blue
or
White
=
Neutral
Brown
or
Black
=
Live / line hot.
5.2.
Install equipment is used on a clean, dry, non-combustible, solid work surface with at
least 300mm suitable clearance all around from other equipment / objects.