• Severe dehydration may lead to low measurement values.
• An abnormally high concentration of vitamin C (ascorbic acid) may
lead to high measurement values.
• The hematocrit (red blood cell volume) level in the blood may
influence the measured results. If the hematocrit level is exceeds
55% (>55%), the measurement results may be too low; however, if
the hematocrit level is below 30% (<30%), the meter may display
unusually high results.
• In case of dialysis the hematocrit level may vary. This may
influence the displayed glucose level.
BLOOD SAMPLING
It is recommended to use the lancing device and disposable lancets
packed together with the meter to prick your finger. If you use them,
please read the following instructions: