IMPORTANT!
•
Read the
AutoSense Test
strip Instructions for Use before using them.
•
Always note the date when the vial is first opened, and use the strips
within 3 months from that date.
•
Use the removed strip immediately and always close the vial carefully
immediately after removing a strip.
•
If the strips are stored below or above room temperature, wait at
least 30 minutes before opening the vial.
Strip insertion and use
The strip must be inserted into the meter’s strip holder with the black side
facing upwards and the square end pointing towards the device. The
meter will switch on automatically upon the insertion of the test strip.
Always ensure that the strip is completely and securely inserted into the
center of the strip holder.
Ensure that enough blood (0.5
μ
l) is applied to the reagent area located
at the tip of the arrowhead-shaped end of the strip. The reagent window