10
Electronics Inc. • (574) 256-5001 • www.electronics-inc.com
Trouble Shooting the Model TSP-M
If the TSP-M Almen Gage indicator display is flashing, or if the gage will not zero, or the resolution needs to be
changed, the following instructions will return the gage back to its factory default settings and remedy these
conditions. Perform every step—one through six—in sequence until the problem is resolved. If more assistance is
needed, please call Electronics Incorporated Customer Service Department at 1-800-832-5653 (USA and Canada)
or (574) 256-5001.
Step One
•
Press M/OFF and the inch/mm button at the same time. "
M1
" will appear on the screen.
•
If the "R" is not on the bottom left side of the screen, press ZERO/ON button until the “R” is on the screen.
•
If the "X1" is not on the bottom center of the screen, press the inch/mm button until “X1” is selected.
Step Two
Press M/OFF button. "
M2
" will appear on the screen and it should read
000.0000
.
•
If “-“ precedes the digits, press the ZERO/ON button to remove it.
•
Press the inch/mm button to move the cursor over to the first non-zero digit to be changed.
•
Press the ZERO/ON button repeatedly to toggle through numbers until “0” is displayed. Repeat for other
digits not at “0.”
Step Three
Press M/OFF button. "
M12
" will appear on the screen and it should read
000.0000
.
•
If “-“ precedes the digits, press the ZERO/ON button to remove it.
•
Press the inch/mm button to move a cursor over to the first non-zero digit to be changed.
•
Press the ZERO/ON button repeatedly to toggle through numbers until “0” is displayed. Repeat for other
digits not at “0.”
Step Four
Press M/OFF button. "
M23
" will appear on the screen and it should read
000.0000.
•
If “-“ precedes the digits, press the ZERO/ON to remove it.
•
Press the inch/mm button to move a cursor over to the first non-zero digit to be changed.
•
Press the ZERO/ON button repeatedly to toggle through numbers until “0” is displayed. Repeat for other
digits not at “0.”
Step Five
•
Press M/OFF button. "
M13
" will appear on the screen. This is the Digital Resolution Mode screen.
•
When using the mm mode – press the inch/mm button repeatedly to select
0.001
.
•
When using the inch mode – press the inch/mm button repeatedly to select
0.0001.
Step Six
•
Press M/OFF button - the gage will be in run mode.
•
Place the flat side of the curved Zero block on top of gage and press the ZERO/ON button.
•
The indicator will read 0.000 mm if in metric mode or 0.0000 if in English mode.
•
The bottom of the screen will read
"R X1 mm" or "R X1 in".
Step Seven
Check the Spindle Starting Position to ensure it is in the preferred range. See the next section for directions on
checking or setting the Spindle Starting Position.
Summary of Contents for TSP-M
Page 1: ...IM0124 2018 04 ...
Page 15: ...Electronics Inc 574 256 5001 www electronics inc com 15 Graph 1 3 26 A to 1 15 Mini A ...
Page 16: ...16 Electronics Inc 574 256 5001 www electronics inc com Graph 2 3 25 A to 1 25 Mini A ...
Page 19: ...Electronics Inc 574 256 5001 www electronics inc com 19 Graph 5 5 95 A to 2 46 Mini A ...
Page 20: ...20 Electronics Inc 574 256 5001 www electronics inc com Graph 6 6 09 A to 2 40 Mini A ...
Page 22: ...22 Electronics Inc 574 256 5001 www electronics inc com Graph 8 12 32 A to 4 93 Mini A ...
Page 23: ...Electronics Inc 574 256 5001 www electronics inc com 23 Graph 9 12 08 A to 5 68 Mini A ...