A993
OTS5000
18
2.3
FUNCTIONAL CHECKOUT
1. Ensure that the unit is installed correctly per section 2.2.
2. Activate power using switch on the back panel. (Indicated by the l side of the rocker switch
pressed in.)
3. Ensure that the LEDs in the front of the blade head are illuminated.
4.
Ensure that the three digit display illuminates, and reads “000”. If the display reads otherwise,
press the ZERO button and ensure that it displays “000”.
5.
Ensure that both bar graph LED’s are at 10 if this is the initial checkout. If this is a yearly
checkup, use the appropriate SPEED buttons to set both displays to 10.
6.
Press the OSCILLATION “+” button repeatedly to advance the display to 15. Ensure that the
display moves up and alternates between blinking and
moving to the next bar. Hold the “+”
key down. Ensure that after one second the display advances rapidly to the top (20).
7.
Press the OSCILLATION “-” button repeatedly to retract the display to 15. Ensure that the
display has moved down and alternates between blinking and moving to the next bar. Hold
the “-” key down. Ensure that after one second the display retracts rapidly to the bottom (0).
8.
Repeat steps 6 and 7 using the ADVANCE “+” and “-“ buttons. When complete use the
appropriate buttons to return both displays to 10.
9.
Press either of the THICKNESS buttons once. Ensure that the three digit display reads “100”
if this is the initial checkout. If this is a yearly checkup, use the appropriate THICKNESS
buttons to set to 100.
10.
Press the THICKNESS “+” button repeatedly. Ensure that the display increments by one with
each press of the button. The first press would cause the display to show the current
thickness if it was previously showing the slice count. Hold the THICKNESS “+” button down.
Ensure that after one second the display advances rapidly.
11.
Press the THICKNESS “-” button repeatedly. Ensure that the display decrements by one with
each press of the button. The first press would cause the display to show the current
thickness if it was previously showing
the slice count. Hold the THICKNESS “-” button down.
Ensure that after one second the display advances rapidly.
12. Using the THICKNESS buttons, set the slice thickness to 500.
13. Press the HOME button. If this is the initial checkout, no movement of the blade arm should
occur. If this is a yearly checkout, the blade arm may return to its home position if it was not
previously returned to it.
14. Press the DOWN BLADE POSITION button once. You should observe the blade arm moving
down. Ensure that the Slice Counter display has incremented by one. Hold the DOWN
button. Ensure that the blade arm moves continuously down. Releasing the DOWN button
stops movement.
15. Repeat #14 with the UP button. Ensure that the blade arm moves up a step amount when
pressed once, and moves up continuously when held down.
16. Press the FORWARD button once. Ensure that the blade begins oscillating and moving
forward. Press the FORWARD button again, ensure that the blade stops all movement.
17. Press the HOME button once. Ensure that the blade arm moves to its full up and back
position.
18. Press the START/STOP button. Ensure that the blade arm steps down, begins oscillating and
moving forward, and that the slice counter increments by 1.