REFERENCE MANUAL
2.1
REFERENCE INFORMATION
2.1.1
PACKAGING
The EMS 5000 is enclosed in a two-piece (bottom and back, top) ABS UL94V0 plastic enclosure . The
overall dimensions are 27cm x 40cm x 21.5cm (10.5" x 16" x 7.5") The top portion is connected to the
interior chassis on the top of the unit, the back panel, and the bottom portion of the case through the
baseplate.
2.1.2
MOUNTING
The EMS 5000 is designed as a stand-alone unit. It is not designed to be rack mounted.
2.1.3
INSPECTION
EMS Modules are factory checked and calibrated but should be carefully inspected upon receipt,
before using, or activating power. If any exterior damage to the shipping carton is noted, the
instrument(s) should be inspected for obvious physical damage. The contents of each package
should be physically checked against the inventory list (sec. 1.3) to determine shortages or errors in
inventory.
2.1.4
POWER CONNECTIONS
The EMS 5000 is powered through an internal 12V power supply. (input:85-265VAC, 50-60Hz, 500mA
max) An international pattern Line Cord (not shown) is ordered separately, and is specified by country
per the catalog number. (See table below for catalog numbers.) The power entry uses double pole
switching and fusing. (2 ea. 5X20, 1/2 amp slo-blo fuses) Contact Technical Services at 215-412-8400
for assistance.
55-AUS Australia
55-CH China
55-DAN Denmark
55-EURO Europe
55-ISR Israel
55-ITA Italy
55-JA Japan
55-SAF South Africa
55-SWI Switzerland
55-UK United Kingdom
55-USA North America
To replace fuses:
1. Remove line cord if applicable
2. Remove fuse drawer located between the line cord entry and power switch by squeezing the
tabs on each end of the drawer, and pulling out.
3. Replace fuses with the correct value (2 ea. 5X20, 1/2 amp slo-blo)
4.
Insert drawer back into slot until tabs “click”.