Overview of the switches and LCD display 

The Primary display of the large LCD shows the result of the test being conducted. At 
the same time a secondary display area shows supporting information e.g. Pass/Fail. 

 

 
 
 
 

 

 
 
 
 
 
 

 
 
 
 
 

 

 
 
 
 
 
 
 
 

 

 

 

 

 

Mains supply wiring and voltage test 

When first connected to a mains supply the AUTOTRIP will automatically conduct a

 

safety test to ensure that the Live, Neutral and Earth conductors are all

 

connected correctly and that the supply voltage is in the acceptable range

 

(195-253V).

 

Please note the polarity LED indicator is not able to determine N-PE supply 

reversal.

 

If all is well the VOLTAGE/POLARITY warning LED will light Green and the

 

supply voltage will be displayed in the primary display area.

 

 

In the event of a problem with either the mains voltage supply or reversed

 

connections the VOLTAGE/POLARITY warning LED will light Red, a warning

 

tone will be sounded and testing will be inhibited.

 

 

Battery Condition 

Indication 

Manual Select Test 

Button 

½

, x1, x5 or Ramp 

Select RCD Rating 

10mA, 30mA, 100mA  

Type AC 

Start Test Button 

Select RCD 100mA 

Type ACS 

 

Hands Free Function 

 

Rotary Selector 

LCD Backlit Display 

Result Recall Button 

 Quick Auto & Full Auto 

Tests 

30mA Type II (AC) 

 

Mains Polarity Check 

Voltage Present Indication 

Summary of Contents for AUTOTRIP

Page 1: ...AUTOTRIP RCD TESTER USER MANUAL...

Page 2: ...ty of the protective earthing conductors and equipotential bonds before carrying out RCD testing Do not operate the meter around explosive gas vapor or dust Do not apply more than the rated voltage 30...

Page 3: ...f and disconnect from any electrical circuits Remove test leads Remove the small crosshead screw on the back of the instrument that retains the battery cover and fit four Alkaline batteries type AA LR...

Page 4: ...n which the tester is primed to automatically start the test as soon as the probes are connected to a circuit thereby leaving your hands free to hold the test probes Socket wiring check To protect bot...

Page 5: ...d set that can be fitted with either prod tips or crocodile clips as required The 3 colour coded prods crocodile clips of the test lead should be connected to the corresponding Line Neutral and Earth...

Page 6: ...indicator is not able to determine N PE supply reversal If all is well the VOLTAGE POLARITY warning LED will light Green and the supply voltage will be displayed in the primary display area In the ev...

Page 7: ...tion time of 40ms when a fault of five times itsrated current is introduced This is referred to as the x5 test For the reasons explained below all of the above tests have to be conducted at both 0 and...

Page 8: ...S When first connected to a mains supply the AUTOTRIP will automatically conduct a safety test to ensure that the Live Neutral and Earth conductors are all connected correctly and that the supply vol...

Page 9: ...on to conduct a test at the x1 setting at 0 The result will be shown as a pass if a 30mA RCD trips within 300ms After displaying the result for a few seconds the tester will switch to the 180 phase po...

Page 10: ...be recalled by using the RECALL button to cycle through the routine The AUTOTRIP will also indicate whether it has passed or failed the RCD requirements 30mA Full Auto Test Turn the rotary selector t...

Page 11: ...CD Trip Time Measurement Range 5ms 1999ms Operating Range Per EN61557 38ms 1999ms Standards To meet 300V CATIII The AUTOTRIP complies fully with the requirements of EN61010 and complies with the secti...

Page 12: ...air and calibration please contact place of purchase Electrical Factory Outlet Pty Ltd PO Box 701 Rochedale South QLD 4123 PH 61 7 3411 0303 E mail sales electricalfactory com au www electricalfactory...