MIN/ MAX Mode:
- Accuracy: same as previous table +0.3% L
- Capture time: 100 ms typ. (every 400 ms)
3.9 Power factor calculation (PF)
With the clamp configured in power measurement (switch on position
W
) and correctly connected (see § 3.8),
perform a short press on the yellow key: the power factor is displayed.
Display range
Measuring range(1)
Accuracy
Resolution
0,20 a
0,49
0,50 a
1,00
5% L +2 pt
1,00
0,01
5% L +2 pt
Characteristics
(1) The display of the power factor is limited to 1.00 If one of the terms of the power factor calculation is outside its
power range, the display of the power factor indicates an indeterminate value
- - - -
.
MIN/ MAX Mode:
-Accuracy: Same for the table below + 1 pt
-Capture time: 100 ms typ. (every 400 ms)
3.10 Frequency Measurement (Hz)
This function is active for measurements V, A, W in AC. For the power function, the frequency measurement is
performed on the voltage signal.
1.
Perform a short pressing on the Hz key, the display shows the frequency of the measured signal.
2.
Pressing again returns to the previously displayed measurement.
Display range
40 Hz
Measuring range(1)
Accuracy
Resolution
10,00 a
39,99 Hz
0,4% L +1 pt
0,01 Hz
400 Hz
4000 Hz
Triggering threshold (2)
5V ou 10A
0,1 Hz
1 Hz
40kHz
40,0 a
399,9 Hz
400 a
3999 Hz
4,00 kHz a
19,99 Hz
10 Hz
1) Below 5 Hz, the display shows
0.0
(2) Below the triggering threshold, the display shows an indeterminate value (- - - -
).
MIN/MAX
- Accuracy: same as in the table above +0.2% of reading with limitation to 5 kHz.
- Capture time: 125 ms typ. every 400 ms
3.11 Phase order indication
This measurement is performed with 2 cables only, sequentially as follows:
1.
Integration of a reference period on one phase L1-L2 for example
2.
Integration of a reference period on one phase L1-L3
3.
Calculation of the time delay between reference and measurement periods enabling determination of the phase
order or phase rotation direction.
7
Characteristics