Manuale di istruzioni/Scheda tecnica

La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.

The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.

La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.

ART. 09/08730-00

PAGINA 5 DI 9

2.2 Button description

•  Button

 

  Switch for meter powering on/off.

•  Button

  

  Switch for turning on/off the backlight 

•  HOLD Button

 

  Switch for reading retention mode. 

•  FUN

 

Button

  It is used for switching among the functions of wind speed  

 

  measurement, area setting and air volume measurement. Long press  

  for three seconds to enable or disable "Auto Power-Off" function. 

•  MAX/MIN Button

 

  Switch maximum/minimum/normal mode, long press to exit.

•  UNIT Button 

  Switch unit, area (m², ft²),

  Wind speed (m/s, km/h, mil/h, ft/m, ft/s, knots), 

  Air volume (CMS, CMM, CFM).  

2.3 Sign definition

      

Indicates Auto Power-Off status

HOLD

    Reading hold state

VEL

      Wind speed measurement state

FLOW

   Air volume measurement state

AREA

    Area setting required by air volume

SLOW

    The current wind speed less than 5m/s

FAST

    The current wind speed greater than 5m/s

MAX

    Displays maximum after entering maximum/minimum mode 

MIN

     Displays minimum after entering maximum/minimum mode 

    

Indicates that the current area setting unit is square meter

ft²

    

Indicates that the current area setting unit is square foot

CMM

     Cubic meters per minute 

CMS

      Cubic meters per second

CFM

      Cubic meters per minute

knots

    Nautical miles per hour, 1850 meters per hour 

ft/s

       Feet per second

ft/m

      Feet per minute

m/s

      Meters per second

Km/h

     Kilometers per hour 

mil/h

     Nautical miles per hour

 

Low battery indicator

3. Specification

The meter should be recalibrated under the condition of 18°C÷28°C, 

relative humidity <75% every year.  

3.1 General

3.1.1   Work height: Maximum 2000 mt.

3.1.2   Work mode: Frequency of wind speed conversion

3.1.3   Display: LCD

3.1.4   Maximum show value: 9999

3.1.5   Sampling time: About 0.4s/time. 

3.1.6   Low battery indicator:            sign displays on LCD. 

3.1.7   Work power: 1×9V 6F22 battery. 

3.1.8   Operation environment: 

 

Relative humidity → 0~85%RH, no moisture condensation

 

Temperature → 0°C~40°C, no moisture condensation

3.1.9   Detector (fan) operation environment: 

 

Relative humidity → 0~95%RH, no moisture condensation. 

 

Temperature → -20°C~80°C, no moisture condensation. 

3.1.10 Storage environment: 

 

Relative humidity → 0~80%RH, no moisture condensation

 

Temperature → -10°C~50°C, no moisture condensation

3.1.11  Dimension: Meter → 165Lx85Wx38H mm. 

3.1.12  Weight: About 200 gr.

3.2 Technique data

Ambient temperature: 23±5°C, Relative humidity: <75%

3.2.1 m/s

Measuring range

Resolution

Accuracy

0.80 ÷ 30.00 m/s

0.01 m/s

±(2.0% r 50 characters)

30.00 ÷ 40.00 m/s

For reference only

3.2.2 km/h

Measuring range

Resolution

Accuracy

1.40 ÷ 108.00 km/h

0.01km/h

±(2.0% r 50 characters)

108.0 ÷ 144.0 km/h

For reference only

3.2.3 ft/s

Measuring range

Resolution

Accuracy

1.30 ÷ 98.50 ft/s

0.01 ft/s

±(2.0% r 50 characters)

98.50 ÷ 131.20 ft/s

For reference only

3.2.4 Knots

Measuring range

Resolution

Accuracy

0.80 ÷ 58.30 knots

0.01 knots

±(2.0% r 50 characters)

58.30÷77.70 knots

For reference only

3.2.5 mil/h

Measuring range

Resolution

Accuracy

0.90 ÷ 67.20 mil/h

0.01 mil/h

±(2.0% r 5 characters)

67.20 ÷ 90.00 mil/h

For reference only

3.2.6 ft/m

Measuring range

Resolution

Accuracy

78 ÷ 5900 ft/m

1 ft/m

±(2.0% r 5 characters)

5900 ÷ 7874 ft/m

For reference only

3.2.7 Air volume unit

CFM

0 - 99990

(Area) 0 - 9.999 ft²

CMM

0 - 99990

(Area) 0 - 9.999 m²

CMS

0  - 9999

(Area) 0 - 9.999 m²

4. Operating guidance

4.1 Power On/Off

Press the “       ” key to turn on or off the anemometer power.  

4.2 Reading hold

In the measurement process, if the reading should be kept, press "HOLD" 

key to lock the reading, and display the HOLD symbol; press it again to 

unlock. 

Nota:

 In the reading hold state, “FUN”, “MAX/MIN” and “UNIT” keys are 

invalid. 

4.3 Backlight

In the measurement process, if the ambient light is too dark to read, you can 

press "      " key to open the backlight. Backlight timer is set to 15 seconds. 

During this period, you can press "      " key again to turn off backlight at 

any time.

Nota: 

•  The luminous body of backlight is LED with large operating current. 

Frequently using backlight will shorten battery life. Do not use backlight 

when unnecessary. 

•  When the battery voltage ≤7V, the "        " (low battery) symbol will show 

on the display. However, in the case of using the backlight, if the battery 

voltage  ≥7V,  the  battery  voltage  drops  because  of  its  larger  operating 

current, low battery symbol may show (when low battery symbol is 

showing, the accuracy of the measurement can’t be guaranteed). At this 

moment, you needn’t replace battery until the low battery symbol display 

again under normal use condition without using backlight.

- +

- +

- +

Summary of Contents for 09/08730-00

Page 1: ...n abbia subito danni durante il trasporto 1 1 2 Se é stato immagazzinato o trasportato in rude condizioni controllare assolutamente che non abbia subito danni prima di iniziare ad usarlo 1 2 Durante l utilizzo 1 2 1 Questo anemometro deve essere utilizzato unicamente nei limiti specificati di umidità e temperatura ambiente 1 2 2 Se nota alcuna anomalia o guasto interromperne subito l utilizzo 1 2 ...

Page 2: ...circa 0 4s tempo 3 1 6 Indicatore batteria scarica appare sul display LCD 3 1 7 Alimentazione 1 batteria di 9V 6F22 3 1 8 Condizioni operative Umidità relativa 0 85 RH senza condensa Temperatura 0 C 40 C senza condensa 3 1 9 Condizioni operative del ventilatore Umidità relativa 0 95 RH senza condensa Temperatura 20 C 80 C senza condensa 3 1 10 Condizioni di stoccaggio Umidità relativa 0 80 RH senz...

Page 3: ...di prova Usare il tasto FUN per entrare nella modalità di misurazione del volume dell aria il simbolo FLOW appare sullo schermo La misura viene fatta con la superficie del ventilatore perpendicolare alla direzione dell aria Nota 1 Se il rilevatore ventola non mira la direzione del vento la misurazione può essere compromessa 2 Per venti constanti il rilevatore ventola ottiene una lettura ottimale q...

Page 4: ... if you use and protect it appropriately 1 1 Preparation 1 1 1 Please check for damage during transportation after receiving the anemometer 1 1 2 If it should be stored and shipped under hard conditions please confirm that whether the meter is damaged 1 2 Usage 1 2 1 The meter should be used in the range of specified ambient temperature and humidity 1 2 2 If you notice any abnormality or failure i...

Page 5: ...tor sign displays on LCD 3 1 7 Work power 1 9V 6F22 battery 3 1 8 Operation environment Relative humidity 0 85 RH no moisture condensation Temperature 0 C 40 C no moisture condensation 3 1 9 Detector fan operation environment Relative humidity 0 95 RH no moisture condensation Temperature 20 C 80 C no moisture condensation 3 1 10 Storage environment Relative humidity 0 80 RH no moisture condensatio...

Page 6: ...tector fan into the test environment Use the FUN key to set the meter to the Air volume measurement mode the FLOW symbol should display on the screen Measurement is done with the fan surface perpendicular to the wind direction Note 1 If the detector fan is not aligned in the wind direction the measurement can be skewed 2 For steady wind the detector fan will get maximum reading when it is aiming a...

Page 7: ...epción del aparato se ruega que controle que no haya sido alterado durante el transporte 1 1 2 Si ha estado almacenado o transportado en condiciones desfavorables controle absolutamente que no haya subido daños antes de empezar a utilizarlo 1 2 Durante el uso 1 2 1 Este anemómetro debe utilizarse únicamente dentro de los límites definidos de humedad y temperatura ambiental 1 2 2 En caso de anomalí...

Page 8: ...ja este símbolo aparece en la pantalla LCD 3 1 7 Alimentación 1 batería de 9V 6F22 3 1 8 Condiciones operativas Humedad relativa 0 85 RH sin condensación Temperatura 0 C 40 C sin condensación 3 1 9 Condiciones operativas del ventilador Humedad relativa 0 95 RH sin condensación Temperatura 20 C 80 C sin condensación 3 1 10 Condiciones de almacenamiento Humedad relativa 0 80 RH sin condensación Temp...

Page 9: ...te a la dirección del viento Nota 1 Si el detector ventilador no apunta hacia la dirección del viento la medición puede ser comprometida 2 Con vientos constantes el detector ventilador obtiene una lectura óptima al ser dirigido hacia la dirección del viento 4 8 Selección de la unidad del flujo volumétrico Cuando se usa el anemómetro para calcular el flujo volumétrico del aire pulsando el botón UNI...

Reviews: