Page 32 of 46
User Manual
ECD-3-nano
Release 1.52
11 Recording the Displacement Signal with an External
Potentiostat
Many of today’s battery testers and potentiostats provide additional analog inputs that may
be used to record sensor signals along with cell current and potential.
In the following, the combination of the ECD-3-nano with a Biologic potentiostat (MPG-2, SP,
VSP and VMP series) is described as an example. The Biologic potentiostats feature two
analog inputs that are used here to record both displacement and temperature.
1.
Connect the 9-pin Sub-D connector of the optional analog output cable to the analog
input of the respective VMP3 channel.
2.
In the Biologic EC-Lab software, load the experiment settings
ECD-3-nano.mps
provided
on the EC-Link installation media. The settings are shown in the External Devices dialog
(see screenshot below; actual settings may differ). Adapt the
Parameter
Settings
of the
charge/ discharge protocol to your particular experiment, if necessary.