8 Further assembly outside the glove box
7.25 Inside the glove box:
The cell is now
assembled and hermetically sealed, and
can be removed from the glove box.
7.26
Hook the assembly into the bracket
and fasten it with the two knurled screws.
3
2
1
7.27
Unlock the locking screw (1) and the
excenter (2). Then turn the micrometer
screw (3) clockwise until the sensor tip is
in the upmost position.
Note:
Not following the above instruction
may result in damage of the membrane,
the glass frit, or even the sensor unit,
when attaching the sensor unit onto the
dilatometer cell.