Page 18 of 46
User Manual
ECD-3-nano
Release 1.52
7 Assembling the cell inside the glove box
After moving the different parts of the disassembled cell body into the glove box, follow the
steps below. Protect yourself and handle the chemicals with care.
7.1 Inside the glove box:
Insert the T-frit
with the smaller side pointing downwards
into the frit flange.
Note:
Make sure that the inside of the frit
flange does not get damaged/ scratched
when inserting the T-frit. Replace the frit
flange as necessary.
7.2 Inside the glove box:
Put a glass fiber
separator (12 mm diameter) on top of the
frit
7.3 Inside the glove box:
Insert the
lithium metal counter electrode.