Appendix
30
Electromagnetic immunity
The RadiForce series has been tested at the following compliance levels according to the
testing requirements for professional healthcare facility environments specified in IEC /
EN60601-1-2.
The customer or the user of the RadiForce series should assure that it is used in such an
environment.
Immunity test
Test level for
professional
healthcare facility
environments
Compliance level
Electromagnetic environment -
Guidance
Electrostatic
discharge (ESD)
IEC / EN61000-4-2
±8 kV contact
discharge
±15 kV air discharge
±8 kV contact
discharge
±15 kV air discharge
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transients / bursts
IEC / EN61000-4-4
±2 kV power lines
±1 kV input / output
lines
±2 kV power lines
±1 kV input / output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surges
IEC / EN61000-4-5
±1 kV line to line
±2 kV line to ground
±1 kV line to line
±2 kV line to ground
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC / EN61000-4-11
0 % U
T
(100 % dip in
U
T
) 0.5 cycles and 1
cycle
70 % U
T
(30 % dip in
U
T
) 25 cycles / 50 Hz
0 % U
T
(100 % dip in
U
T
) 250 cycles / 50
Hz
0 % U
T
(100 % dip in
U
T
) 0.5 cycles and 1
cycle
70 % U
T
(30 % dip in
U
T
) 25 cycles / 50 Hz
0 % U
T
(100 % dip in
U
T
) 250 cycles / 50
Hz
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the RadiForce series requires
continued operation during power
mains interruptions, it is
recommended that the
RadiForce series be powered
from an uninterruptible power
supply or a battery.
Power frequency
magnetic fields
IEC / EN61000-4-8
30 A/m
(50 / 60 Hz)
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
The product should be kept at
least 15 cm away from the
source of power frequency
magnetic fields during use.