Aquameter
™
Series Instruction Manual
10401-00188 Rev M
7.
Select pH7.00. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Calibrating
100%
Press [OK]
When calibration is complete, press the
OK
key then the
ESC
key repeatedly to return to
normal operating mode.
Remove the Probe from the calibration bottle, rinse thoroughly in de-ionised water, shake
off any excess and dry the outer sleeve with a soft cloth.
Now fill a 300mL calibration bottle with fresh pH 4.01 solution and drop the Probe in all the
way. Follow the procedure detailed above, but at step 6, select pH4.01. Wait while the
Meter stabilises and calibrates. When the ‘Calibrating 100%’ screen (shown above) is
displayed, press
OK
then press the
ESC
key repeatedly to get back to the main display.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water, shake off any
excess and dry the outer sleeve with a soft cloth. Dampen the sponge in the storage cap
with storage solution and fit it to the pH/ORP electrode. pH calibration is now complete.
13.5. Errors During Calibration
If a problem occurs during calibration, an error message will be displayed. Refer to
in section 10 for error message handling.
13.6. pH Electrode Efficiency
If the pH electrode becomes worn or clogged, its efficiency and response time can be
reduced. The efficiency of the pH electrode is constantly monitored and in the event of the
efficiency dropping below 85%, ‘ERROR 01’ will be flashed on the bottom line of the
display. If this occurs, or if the pH reading response becomes slow, recondition the
electrode as described below.
1.
Remove the pH or combined pH/ORP electrode from the Probe body
(see
Electrode Removal and Replacement
2.
Rinse with methyl alcohol.
3.
Replace the electrode.
4.
Re-calibrate.
© 2011 Aquaread Ltd.
www.aquaread.co.uk
Page 28 of 64