Move the cursor arrow to the electrode you want to reset, then press the
MR
key.
A confirmation screen will be displayed.
Are you sure you
want to restore the
factory calibration
values? [ESC]=NO
If you are sure, press the
OK
key. If you want to change your mind, press the
ESC
key. If
you press OK, you will see a message that says CAL RESTORED.
Once factory calibration defaults have been restored, you
must
carry out a
full calibration
of the electrode in question.
10.5. Calibration Data Storage
Each Aquaprobe contains its own microprocessor and memory. All calibration data,
including the GLP data, is stored within the Probe’s memory. When a Probe is connected to
a Meter, this data is transferred for display and logging.
This is a major advantage and allows you to use a variety of different Probes with a single
Meter, without the need for re-calibration.