10 EMC Tables
Page 36 / 41
Proximity fields
from RF
wireless
communication
equipment
See the RF
wireless
communicati
on
equipment
table in
"Recommen
ded
minimum
separation
distances"
Complies
minimum separation
distances"
IEC 61000-4-3
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various
healthcare locations where medical equipment and/or systems are used. When
they are used in close proximity to medical equipment and/or systems, the
medical equipment and/or systems’ basic safety and essential performance
may be affected. The
AirPex
has been tested with the immunity test level in the
below table and meet the related requirements of IEC 60601-1-2:2014. The
customer and/or user should help keep a minimum distance between RF
wireless communications equipments and the
AirPex
as recommended below.
Test
frequency
(MHz)
Band
(MHz)
Service
Modulation
Max
power
(W)
Distance
(m)
Immunity
test
level
(V/m)
385
380-
TETRA
Pulse
1.8
0.3
27