background image

11      Statement 

                                                                                                   

Page    40 / 41 

11.Statement 

Service Life 

The service life of AirPex series products is 3 years. 

Maintenance 

MANUFACTURE  will  provide  circuit  diagrams,  component  part  lists, 

descriptions, 

calibration 

instructions 

to 

assist 

to 

SERVICE 

PERSONNEL in parts repair.

 

Disposal 

The package should be recycled. Metal parts of the device are disposed 

as scrap metal. Synthetic materials, electrical components, and printed 

circuit boards are disposed as electrical scrap. The lithium batteries are 

disposed as special refuse. Please deal with them according to the local 

environmental protection laws and regulation. 

Rights 

All  rights  of  modifying  the  product  are  reserved  to  the  manufacturer 

without  further  notice.  The  pictures  are  only  for  reference.  The  final 

interpretation  rights  belong  to  CHANGZHOU  SIFARY  MEDICAL 

TECHNOLOGY  CO.,  LTD.  The  industrial  design,  inner  structure,  etc, 

have claimed for several patents by SIFARY, any copy or fake product 

must take legal responsibilities.

 

 

Summary of Contents for AirPex

Page 1: ...AirPex Apex locator USER MANUAL Changzhou Sifary Medical Technology Co Ltd...

Page 2: ......

Page 3: ...ntended Use 8 3 2 Contraindications 8 4 Installing the AirPex 10 4 1 Connecting file clip lip hook and extension cord 10 4 2 Connecting charge base 11 5 Use Interface 12 6 Operation 13 6 1 Charge 13 6...

Page 4: ...1 Scope of AirPex Page 4 41 1 Scope of AirPex 1 1 Parts Identification 1 2 3 4 6 5 7 8 9 1 Charge Base 2 APEX Locator 3 Lip Hook 4 Touch Probe 5 File Clip 6 Extension Cord 7 Adapter 8 Tester 9 Clip...

Page 5: ...o 6151002 Charge Base 1pcs Part No 6151001 Adapter 1pcs Part No 6516001 Lip Hook 2pcs Part No 6072002 Touch Probe 1pcs Part No 6101001 File Clip 2pcs Part No 6151003 Extension Cord 2pcs Part No 615100...

Page 6: ...eration may lead to hazards for the product or the user patient NOTE Additional information explanation of operation and performance Serial number Catalogue number Manufacturer Date of manufacture LOT...

Page 7: ...134 C Can be autoclaved up to a maximum temperature of 134 Celsius Authorized Representative in the European Community 55 C 20 C Temperature limitation 80 20 Relative humidity 106kPa 70kPa Atmospheric...

Page 8: ...ther electrical equipment and has been cautioned against the use of small electrical appliances such as electric shavers hair dryers etc Safety and effectiveness have not been established in pregnant...

Page 9: ...s such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the AirPex including cables specified by the manufacturer Otherwise degradation of the perfo...

Page 10: ...rd and lip hook to APEX locator as shown in the picture Also use both extension cords according to actual situation NOTE Please use the original file clip and lip hook that manufactured by Sifary Beca...

Page 11: ...TE Only the original adapter could be used Put the APEX locator in the groove in the middle of the charge base The power LED on charge base will light up And the interface of APEX locator will light u...

Page 12: ...cle the volume through the minor to the maximum Setting the reference point During standby state press to set the reference point between 0 1 Seven points can be selected circularly The cursor flashin...

Page 13: ...Operation 6 1 Charge Display the present remaining amount of the battery Less than 15 remains please charge NOTE If the power if less than 15 the device must be recharged within 30 days otherwise the...

Page 14: ...Fully charged will take about 4 5 hours depending on residual battery power and battery state It can be recharged 300 500 times depending on the operating conditions of the device WARNING Do not chan...

Page 15: ...measuring bar on the screen flashes at point 02 03 or 04 Recommend to test the APEX locator with tester once a week NOTE If the measurements are not expected check that the tester is connected proper...

Page 16: ...hook and extension cord according to chapter 5 1 Finally touch the lip hook with exposed metal position on file clip the measurement on the screen should be shown as 2 NOTE If the measurement shown i...

Page 17: ...the file clip and the root canal file handle are basically perpendicular otherwise the chuck of the file clip will be easily damaged This equipment does not include the root canal file We need to buy...

Page 18: ...void causing wrong measurement results APEX locator should be fixed in patient s collar with the clip NOTE To prevent measurement errors caused by conduction between the gums or adjacent root canals d...

Page 19: ...ed by the APEX locator becomes shorter When displayed as shown in figure 3 it means that it reaches the tip of the tooth root The measured value is 00 which is the actual length of the root canal to b...

Page 20: ...al entry WARNING During measuring enter slowly to prevent piercing the tip of the root The APEX locator is used to determine the working length of the root canal In clinical use it must be combined wi...

Page 21: ...ning If blood spills from the root opening and contacts the gums it will cause leakage of electricity which cannot be accurately measured Wait for the bleeding to stop completely Clean the root canal...

Page 22: ...ed to isolate the gingival tissue Branch Crack The crack tooth Leakage through branch of the root canal Broken teeth can cause electrical leakage and cannot be accurately measured Branch tubes can als...

Page 23: ...e crown so that the file will not touch the mental prosthesis before taking a measurement Debris Pulp Cutting debris on tooth Pulp inside canal Remove all cutting debris on the tooth Remove all the pu...

Page 24: ...quite close to the apex In this case try to moisten the canal with oxydol or saline Difference measuring result between Apex locator reading and Radiography Sometimes the reading of the apex locator...

Page 25: ...hat the actual apex of the root canal is not the same as the anatomic end In fact the apical foramen is located at the coronal end in this case X ray may indicate that the file needle has not reached...

Page 26: ...eaned and sterilized before every use Pre cleaning Use a soft brush or moist cloth to remove visible contaminations then wipe all the surfaces with a cloth lightly moistened with Ethanol for disinfect...

Page 27: ...ion 10 minutes Storage Keep the components in sterilization packaging in a dry and clean environment WARNING Comply with your national guidelines standards and requirements for cleaning disinfection a...

Page 28: ...all the surfaces with a cloth lightly moistened with Ethanol for Disinfection Ethanol 70 to 80 vol at least 2min repeat for 5 times NOTE Do not use anything except Ethanol for Disinfection Ethanol 70...

Page 29: ...distributor Problem Cause Solution The power is not turned on The battery is flat Charge the battery Press the power switch too short time Long press the power switch No charge indicator flash on hand...

Page 30: ...gy Co Ltd Model AirPex Dimensions 13cm x 11cm x 8cm 1cm Outer box Weight 0 35kg 10 Power supply Lithium ion battery 3 7V 120mAh 10 Charger power supply AC 100 240 V 10 Charger power output 5V 1A Frequ...

Page 31: ...ting conditions Use in enclosed spaces Ambient temperature 5 C 40 C Relative humidity 80 Operating altitude 3000m above sea level Atmospheric pressure 70kPa 106kPa Transport and storage conditions Amb...

Page 32: ...emissions CISPR 11 Group 1 The AirPex uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...

Page 33: ...act 2 kV 4 kV 8 kV 15 kV air 8 kV contact 2 kV 4 kV 8 kV 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least...

Page 34: ...should be that of a typical commercial or hospital environment If the user of devices require continued operation during power mains interruptions it is recommended that devices be powered form an uni...

Page 35: ...ducted dis turbances induced by RF fields IEC 61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM bands be tween 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 3 V Portable and mobile RF...

Page 36: ...or systems are used When they are used in close proximity to medical equipment and or systems the medical equipment and or systems basic safety and essential performance may be affected The AirPex has...

Page 37: ...710 704 787 LTE Band 13 17 Pulse modulation 217Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 90 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18Hz 2 0 3 28 870 930 1720 1700 1990 GSM 1800 CDMA 19...

Page 38: ...785 WARNING 1 Use of accessories and cables other than those specified or provided by the manufacturer of AirPex could result in increased electromagnetic emissions or decreased electromagnetic immuni...

Page 39: ...AirPex adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary AirPex and the other equipment should be observed to verify...

Page 40: ...rical components and printed circuit boards are disposed as electrical scrap The lithium batteries are disposed as special refuse Please deal with them according to the local environmental protection...

Page 41: ...rict Changzhou City 213000 Jiangsu China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www eighteeth com Llins Service Consulting GmbH Tel 49 175 4870819 Add Obere Seegasse 34 2 6...

Reviews: