11.2 Replace the Fuse
Replace the fuse is to replace the power adapter.
Attention!
1. The fuse is inside the power adapter. Fuse shall be replaced by qualifi ed service personnel who is approved
by EICKEMEYER®.
2. Before replacing the fuse, please contact EICKEMEYER®, replace the fuse under the guidance of EICKEMEYER®.
3. Before replacing the fuse, you must disconnect the mains supply.
4. Fuse Type: T3.15AH250VAC
11.3 Replacement of Power Supply Cord
Before replacing the power supply cord, please contact EICKEMEYER®; replace the power supply cord under the guidance of
EICKEMEYER®. Please use the power supply cord provided by EICKEMEYER®.
11.4 Troubleshooting
To ensure normal operation, users are recommended to prepare a proper maintenance and regular examination plan to
regularly check on product safety performance. If any abnormity occurs, contact EICKEMEYER® for support.
If the following problems occur on starting up the machine, try to make corrections following the method in the table. If the
problem remains unsolved, contact EICKEMEYER® for support.
Trouble
Correction
Power supply indicator is off and no screen display is
present when starting the machine.
1. Check power supply.
2. Check power cable and plug.
3. Check power adapter.
Character and gray scale are displayed, but no ultrasonic
image on the screen.
Probe is not properly connected. Turn off the power and
reconnect the probe.
Intermittent stripe, snow, or far-fi eld interference appears on
screen.
1. Check power supply (spark interference present).
2. Check environment (source of interference around the
machine, such as electric motor, ultrasonic atomizer,
automobile, computer or other interference).
3. Check power plug / socket of the instrument or probe
connectors. They shall be properly contacted.
Image display is not clear.
1. Adjust the total gain, near fi eld gain, far fi eld gain.
2. Adjust the backlight, brightness and contrast level.
Control panel malfunction
Restart the system by turning off the main unit power.