3
INSTALLATION GUIDE
HAICO TELGTE II EQUINE OPERATING TABLE
T +49 7461 96 580 0 | F +49 7461 96 580 90 | [email protected] | www.eickemeyer.com
Using the Battery Charger
This automatic charger is designed for 90 to 270 V AC and 50 to 60 Hz.
In order to start the charge the battery terminal voltage must be over 1 V.
In case the terminal voltage is under 1 V the battery may be in short and it should be replaced.
Telgte series tables
Pro3 tables with magnet plug
Note the hanging way
The built in micro controller controls the charging process.
Yellow LED on
Charge in progress
Max power:
For max power ON – disconnect and connect again the wall plug
Time: 0 – 12 h
Both LEDs on
Battery almost full
14.4 V charge
Time: 4 h
Green LED on
Charge ready
Maintenance charge 13.7 V
No time limit
Attention!
• Keep the magnetic contacts clean from metal crumbs or connection between the charger and the battery may
be cut off.
•
As a safety measure, the charger should be mounted on a wall at a height of about 1.5 m.
The LEDs here