All-in-1 Operations Manual
Page 44
www.ehealthcareinc.com
by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the All-in-1 is used exceeds the applicable RF compliance
level above All-in-1 care should be observed to verify normal operation. If abnormal performance
is observed additional measures may be necessary such
as re‐or
ienting or relocating All-in-1
Health Monitoring.
Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3V/m.