All-in-1 Operations Manual
Page 21
www.ehealthcareinc.com
Figure 10:
Connection of Electrode on
Limbs
When attaching electrode to limbs, follow these simple guidelines:
•
Ensure the electrode is clean.
•
Organize lead wires and do not twist it. Make sure electrode connector and electrode are in solid
connection;
•
Clean skin surface which has contact with electrode by ethyl alcohol;
•
Apply a little conductive paste on skin;
•
Apply thin layer of conductive paste on electrode surface;
•
Clamp skin applied with conductive paste by electrode.
Placement of electrode on chest