3-6
3.2 Signal Conditioning Module Setup
A/E Signal Conditioning Module, E8303
The
polarity
of the input and the air amplification gain must be setup using jumpers on
the module circuit board. The polarity of the input determines whether the input reads
positive or negative as the air nozzles are closed off. The air amplification gain is set to
low, medium, or high depending on the range to be used with the air tooling. See the
“
Advance Operation
” section of the manual for additional information of module setup.
The A/E module contains a MAG and ZERO adjustment knob for input setup. See
section the "Input Setup" section below for A/E input set up procedures.
LVDT Module, E8302
The
polarity
of inputs A and B must be setup using switches on the module circuit board.
The polarity setting determines whether the input reads positive or negative when the
LVDT
tip is depressed. See the “
Advance Operation
” section of the manual for
additional information of module setup.
The LVDT module contains a INPUT selection knob. The INPUT selection knob can be
set to A, B, or AB. When set to A the reading of only input A will be displayed on the
bargraph and the alphanumeric display. When set to B the reading of only input B will
be displayed on the bargraph and the alphanumeric display. When set to AB the result of
input A + input B will be displayed.
The LVDT module also contains a ZERO adjustment knob for inputs A and B, a MAG
adjustment screw for the A and B inputs, and a MAG adjustment screw for inputs C and
D bussed in from a separate column or columns. See section the input setup section
below for LVDT input setup procedures.
Power to the column must be turned off prior to installing or removing a signal
conditioning module.
Summary of Contents for Trendsetter II
Page 2: ...rend Setter II User s Manual T T Edmunds GAGES...
Page 15: ...2 3 2 3 Overall E8300Unit Figure 2 1 Trendsetter II Basic Unit...
Page 17: ...2 5 Figure 2 2 Front Panel...
Page 19: ...2 7 Figure 2 3 Rear Panel...
Page 24: ...2 12 Figure 2 7 E8302 2 Channel LVDT Signal Conditioning Module...
Page 26: ...2 14 Figure 2 8 E8303 1 Channel AE Signal Conditioning Module...
Page 31: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Page 53: ...4 7 Figure 4 5 A E Block...
Page 55: ...4 9 Figure 4 6a LVDT Module E8302 current version...
Page 59: ...4 13 Output Jumpers Input Jumpers Input Type Jumpers Figure 4 8 Input Output Jumpers...
Page 65: ...4 19 J10 J3 J4 Figure 4 10a Attenuation Jumpers current revision...