4-15
Input Type Jumpers(J6, J7)
If a signal from an air to electric module is to be bussed in as input C, variable
magnification, jumper J6 must be set to "A". Set jumper J7 to "A" if the air to electric
signal is bussed in as input D, variable magnification. If a signal from an electronic gage
is to be bussed in as input C, variable magnification, jumper J6 must be set to "E". Set
jumper J7 to "E" if the signal is bussed in as input D, variable magnification.
Gain Ratios
Jumper Setting
Gain Ratio
E
2x - .025x
A
2x - .002x