4.5
Exchange of the Rack Systems of KS 15, TiMix 2 and TiMix 5
Compact Shaker KS 15:
The KS 15 shakers can be delivered with 2 alternative rack systems:
Combifix KS (No. 0052 071) with 3 clamping strips h or universal tray (No. 0051 471)
for spring clamps (see Accessories/ Rack Systems). If you wish to change the rack
system, proceed as follows:
For mounting the universal tray, the Combifix KS must be removed. Remove the
rubber mat and loosen the flat headed screws of the rack system. Remove rack
system.
Fasten the universal tray with the flat headed screws and the distances which are
delivered with the universal tray.
To mount the Combifix KS, proceed in reverse order.
Microplate Shaker TiMix 2:
As an alternative to the integrated standard rack for 4 microplates the shaking plate
of the KM 2 shaker (No. 0052 072) can also be mounted. Loosen the 3 flat headed
screws (M6x16) and remove the rack system. Fasten the shaking plate to the
respective drillings with the same screws and put the rubber mat on the shaking
plate.
Microplate Shaker TiMix 5:
The TiMix 5 shakers can be combined with different rack systems for microplates or
with the Combifix KS (No. 0052 071) or the universal tray KS (No. 0051 471) (see
Accessories / Rack Systems). These rack systems can be exchanged without
problems.
Loosen the flat headed screws of the rack system or tray and fasten the required
rack system instead.
4.6
Fastening Multi-Storey Rack Systems on Shakers WS 10 / TL 10,
SM 30, VKS 75
Tilt Shaker WS 10 / 3-Dimensional Shaker TL 10
The multi-storey rack system WS/TL (No. 0051 247) is fastened with screws to the
corresponding drillings at the back of the basic tray. Place the short black rails on the
inside of the basic tray. Then fasten the knurled screws to the tray and the holes in
the black rails, then tighten the screws.The additional trays can be inserted in any of
the guides of the rack system.
Edmund Bühler GmbH ALLGEM\BED.SMG/BEDSME/TXTE.DOC – 01/08