Using the Field Unit
PhaseTrakker
AP30
26
UNRESOLVED
Records have no
Reference Unit
or Reference
Offset
Phase Shift set-
ting when read-
ing was stored
The Field Unit will continuously attempt to re-establish connection with the config-
ured Reference Unit(s). When either the Primary or Secondary Reference Unit
becomes available, the Field Unit will automatically connect and resolve any
UNRESOLVED Records. Resolving takes place in the background and does not
interrupt any phase readings being taken. The AP30 Reference Unit can hold up
to one month of data, so the Field Unit can resolve UNRESOLVED Records taken
up to one month ago.
When Record is
resolved, Phase
ID and Phase
Angle are stored
Resolving of
UNRESOLVED
records takes
place in the
background
Note: If the UNRESOLVED records are not automatically resolved
when a Reference Unit comes back online, turn the Field Unit power
off and then on again.
Stored Phase
Shift is used to
resolve record
Reference Unit
name and Ref-
erence Offset
are stored